ResMed ApneaLink manual Ordering Information & Product Codes

Page 4

TECHNICAL SPECIFICATIONS APNEALINK AND PULSE OXIMETER

Signal Recording

Breathing sounds

Respiratory flow

Blood oxygen saturation

Pulse

Battery voltage

Sampling Rates for the Channels

Respiratory flow/breathing sounds: 100 Hz

Saturation: 1 Hz

Pulse: 1 Hz

Battery: 1 Hz

Signal Processing

Signal recording: 20 Bit

Signal storage: 16 Bit

Internal Memory

Storage capacity: 15 MB

Recording period: 8 hours minimum

Power Supply to Recorder

2 NiMH rechargeable batteries: Mignon/AA/1.2V/ at least 2.1 Ah or

2 batteries: LR 6/Mignon/AA/1.5V/at least 2.1 Ah

Weight

Recorder (without rechargeable batteries or batteries): Approximately 50 g (1.8 oz)

Pulse Oximeter: Approximately 30 g (1.1 oz)

Operating Conditions

Temperature: 20°C to 40°C (68°F to 104°F)

Humidity: 10% to 90% RH (non condensing)

Shipment/Storage Conditions*

Temperature: -20°C to +50°C (-4°F to +122°F)

Electromagnetic Compatibility

Product complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601- 1-2, for residential, commercial and light industry environments. For further details, see “Guidance and Manufacturer’s Declaration – Electromagnetic Emissions and Immunity” as follows.

Guidance and Manufacturer’s Declaration – Electromagnetic Emissions and Immunity

The ApneaLink is intended for use in the electromagnetic environment specified below. The customer or user of the ApneaLink should ensure that the system is used only in such an environment.

Emissions test

Compliance

Electromagnetic environment—guidance

RF emissions CISPR11

Group 1

The ApneaLink uses RF energy only for its internal function.

 

 

Therefore, its RF emissions are very low and are unlikely to cause

 

 

any interference in nearby electronic equipment

 

 

 

RF emissions CISPR11

Class B

The ApneaLink is suitable for use in all establishments, including

Harmonic Emissions

domestic establishments and those directly connected to the

public low-voltage network that supplies buildings used for

IEC 61000-3-2

 

 

domestic purposes.

Voltage fluctuations/

 

flicker emissions

 

 

IEC 61000-3-3

 

 

 

 

 

Medical Electrical Equipment needs special precautions with respect to EMC and needs to be installed and put into service according to EMC information provided in this document.

Warnings: The ApneaLink should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the ApneaLink should be observed to verify normal, operation in the configuration in which it will be used. The use of accessories other than those specified in this manual is not recommended. They may result in increased emissions or decreased immunity of the ApneaLink.

ORDERING INFORMATION & PRODUCT CODES

Humidity: 10% to 90% RH

*Recorder without rechargeable batteries or batteries

Operating/Storage Air Pressure

800 hPa to 1060 hPa

Effective Range

Flow sensor: -10 hPa to +10 hPa

SpO2: +/- 2 digits

Pulse: +/- 3 digits

Interfaces

Nasal pressure cannula: Luer connection

Pulse oximeter: 3-pin Binder plug

Computer: Fullspeed USB 1.1

Dimensions

Recorder (length x width x height): 125 x 60 x 30 mm (4.9" x 2.4" x 1.2")

Pulse oximeter: (length x width x height): 53 x 20 x 15 mm (2.1" x 0.8" x 0.6")

ApneaLink

US and Latin America

 

22302

Canada

 

22303

1

Apnealink device

1

Program CD

3

Nasal Cannulas

2

AA Batteries

1

Reusable Belt

1

Carrying Case

1

Quick Software Setup Guide

1

USB download cable

 

 

 

ApneaLink Oximetry Accessories Kit

US and Latin America

 

22304

Canada

 

22308

1

Nonin XPOD oximeter

 

 

1

Sensor clip

 

 

3

Disposable Sensors

 

 

Optional Accessories and Disposables (US, Latin America and Canada)

Nasal Cannulas (25/pk)

70388

Nasal/Oxygen Cannula (25/pk)

70319

Belt, reusable

629052

Belt, Single Use (24/pk)

70406

Sensor, Oximeter-Single Use

70412

Sensor, Oximeter (Flex Wrap)-Reusable

1431002

Tape, Sensor (for Flex Wrap Sensor), (25/pk) 70276

Sensor, Oximeter (Soft Sensor)-Reusable

70413

ResMed Corp Poway, CA, USA +1 858 746 2400 or 1 800 424 0737 (toll free), ResMed Ltd Bella Vista, NSW, Australia +61 (2) 8884 1000 or 1 800 658 189 (toll free). Offices in Austria, Brazil, Finland, France, Germany, Hong Kong, Japan, Malaysia, Netherlands, New Zealand, Singapore, Spain, Sweden, Switzerland, United Kingdom (see website for details).

ApneaLink is a trademark of ResMed Ltd. and is registered in U.S. Patent and Trademark Office. ©2006 ResMed. 1011262/1 06 10

Global leaders in sleep and respiratory medicine www.resmed.com

Image 4
Contents ApneaLink Why use ApneaLink? How does ApneaLink work?ApneaLink with oximetry Ordering Information & Product Codes