Maintenance and storage

• For cleaning the monitor only use a soft, lightly

moistened cloth.

Do not use petroleum spirits, thinners or similar

solvents!

• Stains on the cuff can be carefully removed using

a moist cloth and soapy water.

Do not wash the cuff or make it wet!

Technical data

English

This device fulfils the provisions of the EC directive 93/42/EEC (Medical Device Directive). This blood pressure monitor is designed according to the european standard EN1060, Non-invasive sphygmomanometers Part 1: General requirements and Part 3: Supplementary requirements for elec- tromechanical blood pressure measuring systems.

• Protect your monitor against vibrations and do

not leave it in a place where temperatures are

extremely low (below -20 °C) or extremely high

(above 60 °C) or with high humidity (+ 85%).

• Do not use your monitor at very low temperatu-

res (below 10 °C) or very high temperatures

(above 40 °C).

• Do not fold the cuff or air tube together too

tightly and do not crease them.

• Do not carry out repairs of any kind yourself.

If a defect occurs, consult the OMRON

Customer Service or seek the advice of your

surgical appliance stockist or pharmacist.

• The well confirmed accuracy of this blood

pressure monitor is designed to last a long life.

• In order to increase the lifetime of the battery

take it out when you store the device for a

longer period (2 months).

• It is generally recommended to reconfirm the

function and accuracy every 2 years by your

OMRON distributor (address see attached

literature).

Model:

OMRON M1

Display:

LCD digital display

Measurement Range:

Pressure: 0 to 280 mmHg

 

Pulse: 40 – 200/min.

Accuracy/Calibration:

Pressure: ±3 mmHg or 2% of reading

 

Pulse: ±5% of reading

Inflation:

Manual by inflation bulb

Deflation:

Automatic pressure release valve

Rapid Pressure Release:

Push button valve

Pressure Detection:

Capacitive Type Pressure Transducer

Power Source:

4 alkaline batteries (type LR6 1.5 V)

Battery Life:

Capacity of a new batteries is approx.

 

1000 readings

Operating Temperature/

10°C to 40°C

Humidity:

30 to 85% RH

Storage Temperature:

-20°C to 60°C

Console Weight:

Approximately 170 g (including battery)

Outer Dimensions:

Approximately 12.5 (W) x 8.5 (D) x 6.0 (H) cm

Cuff Range:

For arms 22 to 33 cm circumference

Accessories:

Cuff, Type M (CR13); inflation system,

 

instruction manual

Note: These specifications, to improve performance, are subject to change without notice

Manufacturer:

OMRON Matsusaka Co. Ltd., Japan

10

11

Image 6
Contents Semi-automatic blood pressure monitor Contents Dear Customer Important instructions for obtaining meaningful readingsMeasuring your blood pressure ➋ Switching on the unitYour blood pressure monitor the benefits ➊ Prepare the unit➌ Fitting the cuff and Start ➍ The measurement startsHints Failure, causes and rectificationFailure and their possible causes Rectification Do not use petroleum spirits, thinners or similar Maintenance and storageTechnical data SolventsOmron spare parts