✓ TROUBLESHOOTING
Problem | Probable cause | Corrective action | ||
|
|
| ||
Display is blank, | Batteries are drained. Replace with new batteries. | |||
even when the |
|
| ||
Battery terminals are | Place the battery with | |||
power is turned | ||||
on. | not in the correct | negative and positive ter- | ||
|
| position. | minals matching those | |
|
|
| indicated in the battery | |
|
|
| compartment. | |
|
|
|
| |
The device will | The cuff is not fas- | Fasten the cuff correctly. | ||
not measure. | tened properly. |
| ||
Readings are too |
|
|
| |
| You moved your arm | Make sure you remain | ||
high or too low. | ||||
|
| or body during the | very still and quiet during | |
|
| measurement. | the measurement. | |
|
|
|
| |
|
| The position of your | Position your wrist so it | |
|
| wrist is not correct. | is level with your heart, | |
|
|
| using an arm rest if | |
|
|
| needed. | |
|
|
|
| |
Other | The blood pressure | See Pg | ||
|
| reading is different | Measure Blood Pressure | |
|
| from that measured | at Home?” | |
|
| at a clinic or doctor’s |
| |
|
| office. |
|
Remove the batteries. Place back in correctly, and try the measurement again.
NOTE: If the actions described above do not solve the problem, contact A&D Medical or Auto Control Medical at the numbers listed at the bottom of these pages. Do not attempt to repair the device yourself.
LifeSource Health Line |
| SPECIFICATIONS |
| ✓ |
|
|
|
|
Model |
| ||
Type | .Oscillometric |
| |
Display | .Digital, | ||
. . . . . . . . . . . . . . . . . . . . . . | .Pressure/Pulse displayed simultaneously | ||
Memory | .30 readings |
| |
Measurement range | .Pressure: 20 mmHg to 280 mmHg | ||
. . . . . . . . . . . . . . . . . . . . . . | .Pulse: 40 pulses to 199 pulses/minute | ||
Accuracy | .Pressure: +3 mmHg or 2%, whichever | ||
. . . . . . . . . . . . . . . . . . . . . . | .is greater, Pulse: +5% |
| |
Pressurization | .Automatic, using micropump |
| |
Depressurization | .Constant air |
| |
Power source | .2 type AAA (1.5 volt) alkaline batteries | ||
. . . . . . . . . . . . . . . . . . . . . . | .(included) |
| |
Battery life | .Approximately 400 measurements | ||
Operating environment . . . | .50ºF to 104ºF (10ºC to 40ºC) |
| |
| Less than 95% relative humidity |
| |
Storage environment | .14ºF to 140ºF |
| |
| Less than 95% relative humidity |
| |
Wrist Circumference Range | 5.3 – 8.5”(13.5 – 21.5 cm), |
| |
Dimensions | .Depth: 0.8” (21 mm), Width: 2.5” | ||
| (63 mm), Height: 2.2” (56 mm), |
| |
Dimensions with case | .Length: 4.0” (102 mm), Width: 3.0” | ||
| (76 mm), Height: 3.3” (83 mm), |
| |
| approximately |
| |
Weight | .2.9 oz. (82 g) without batteries |
|