TROUBLESHOOTING
Problem | Probable cause | Corrective action | |
|
|
| |
Nothing appears in | Battery is drained. | Replace battery with new one. | |
the display, after |
| Alkaline battery | |
pressing the START |
| recommended. | |
button. |
|
| |
Battery terminals are not in | Reinstall the battery with neg- | ||
| |||
| the correct position. | ative and positive terminals | |
|
| matching those indicated in | |
|
| the battery compartment. | |
The cuff does not | Battery voltage is too low. | Replace battery with new | |
Low battery symbol blinks. | ones. Alkaline battery | ||
deflate properly | |||
| [If the battery is drained | recommended. | |
| completely, the mark does |
| |
| not appear.] |
| |
The unit does not | The cuff is not fastened | Fasten the cuff correctly. | |
measure. Readings | properly. |
| |
are too high or too |
|
| |
You moved your arm or | Make sure you remain very | ||
low. | |||
| body during the | still and quiet during the | |
| measurement. | measurement. | |
| The cuff position is not | Sit comfortably and still. | |
| correct. | Make sure the cuff is at the | |
|
| same level as your heart. | |
| You are using the wrong | See Pg. | |
| size cuff. | Correct Cuff." | |
| Taking too many readings | Relax five minutes before each | |
| on the same arm in a short | measurement. | |
| period of time. |
| |
The value is differ- | The healthcare professional | See Pg. | |
ent from that meas- | may be using a different | Why Your Readings Are | |
ured at a clinic or | sized cuff. | Different Between Home and | |
doctor’s office. |
| the Doctor’s Office." | |
| Your measurements may be | See Pg. | |
| elevated by white coat | Blood Pressure at Home." | |
| hypertension |
| |
|
|
|
NOTE: If the actions described above do not solve the problem, call
LifeSource Health Line |