5.Handling Errors and Problems
5. Handling Errors and Problems
5.1Error Messages
Error Display | Cause | Remedy |
|
|
|
| Cuff is under inflated. | Carefully read and repeat the |
| steps listed under section 3.4. | |
|
| |
|
|
|
|
| Repeat measurement. Remain |
| Movement during | still and do not talk during |
| measurement. | measurement. |
|
| Refer to section 3.3. |
|
|
|
| Air plug disconnected. | Insert the air plug securely. |
| Refer to section 3.2. | |
|
| |
|
|
|
| Arm cuff not applied | Apply the arm cuff correctly. |
| correctly. | Refer to section 3.2. |
|
|
|
| Clothing is interfering | Remove any clothing |
| interfering with the arm cuff. | |
| with the arm cuff. | |
| Refer to section 3.2. | |
|
| |
|
|
|
| Air is leaking from the | Replace cuff with new one. |
| arm cuff. | Refer to Chapter 7. |
|
|
|
| The arm cuff was | Do not inflate the arm cuff |
| inflated above 299 | |
| above 299 mmHg. | |
| mmHg when inflating | |
| Refer to section 3.4. | |
| the cuff manually. | |
|
| |
|
|
|