4.Troubleshooting and Maintenance

4.1The Icons and Error Messages

Error Display

Cause

Remedy

 

 

 

 

 

Remove the arm cuff. Wait 2-3 minutes and then

 

Irregular or weak pulses are detected.

take another measurement.

 

Repeat the steps in section 3.3. If this error

 

 

 

 

continues to appear, contact your doctor.

 

 

 

 

Movement during measurement.

Carefully read and repeat the steps in section 3.3.

 

Cuff is not applied correctly.

Apply the arm cuff correctly. Refer to section 3.1.

 

 

 

 

 

You should replace them with new ones ahead of

Blink

The batteries are low.

time.

 

 

Refer to section 2.1.

 

 

 

Lit

The batteries are exhausted.

You should replace them with new ones at once.

Refer to section 2.1.

 

 

14

Image 16
Contents Model M3W Contents Important Safety Information General UsageImportant Safety Information Overview Main unitInstalling/Replacingthe Batteries PreparationSetting the Buzzer/ Date and Time Press the O/I Start button to store the settingUsing the Unit Applying the Arm CuffHow to Sit Correctly Cuff wrapping guideTaking a Reading Complete If your systolic pressure is Using the Unit Using the Unit Using the Memory Function Press the button, while the average value is displayed Troubleshooting and Maintenance Icons and Error MessagesTroubleshooting and Maintenance Troubleshooting Problem Cause Remedy