Omega Engineering M1982/0902 Flush Mount Front Flange and U-Clamp, Installing the Bulb

Page 3

VA Series Vapor Actuated Dial Thermometer

Flush Mount (Front Flange and U-Clamp)

1.Make a circular cutout in the panel slightly larger than the case.

2.Drill three (3) mounting holes for the Front Flange Model.

3.From the front of the panel, pass the bulb and capillary through the cutout.

4.Affix the head to the panel:

4a. U-clamp Model: Tighten U-clamp(s) against the back of your panel.

4b. Front Flange Model: Screw the flange to the front of your panel, using the screw holes that you drilled.

Installing the Bulb

N O T E

If the bulb is installed more than 10 feet above or below the dial, it may be necessary to recalibrate the instrument.

3

Image 3
Contents Unpacking Instructions Installing the Vapor Actuated ThermometerGeneral Description Available ModelsCase and Mounting Dimensions Installing the Indicating HeadSurface Mount Rear Flange Flush Mount Front Flange and U-Clamp Installing the BulbUnion Connections Standard Bulb Dimensions inchesOperation ThermowellsBendable Extensions Care of the Vapor Actuated Thermometer RecalibrationSpecifications WARRANTY/DISCLAIMER