Omega Engineering M1982/0902 instruction sheet Standard Bulb Dimensions inches, Union Connections

Page 4

VA Series Vapor Actuated Dial Thermometer

Standard Bulb Dimensions (inches)

Immerse the thermometer stem at least A inches (or entire stem, whichever is less) in the medium being measured, for best reading.

Union Connections

C A U T I O N

Tighten the thermometer to its fitting using a wrench applied to the hex or flats above the threads. Never tighten by hand.

1.Unscrew and remove the 12" NPT fitting, using a 78" wrench.

2.Screw the 12" NPT fitting into the threaded process opening.

3.Insert the bulb through the 12" NPT fitting. Tighten it in place by screwing the jam nut into the top of the 12" NPT fitting with an 1116" wrench.

4

Image 4
Contents Installing the Vapor Actuated Thermometer General DescriptionAvailable Models Unpacking InstructionsInstalling the Indicating Head Case and Mounting DimensionsSurface Mount Rear Flange Installing the Bulb Flush Mount Front Flange and U-ClampStandard Bulb Dimensions inches Union ConnectionsThermowells OperationBendable Extensions Recalibration Care of the Vapor Actuated ThermometerSpecifications WARRANTY/DISCLAIMER