Omega Engineering HH504 manual Temperature VS Resistance TABLEITS90

Page 12

TEMPERATURE VS RESISTANCE TABLE(ITS90)

°C

Pt385

Pt3926

 

 

 

-200°C

18.521 W

16.996W

 

 

 

-100°C

60.256 W

59.479W

 

 

 

0°C

100.000 W

100.000W

 

 

 

100°C

138.505 W

139.272W

 

 

 

200°C

175.856 W

177.362W

 

 

 

300°C

212.052 W

214.275W

 

 

 

400°C

247.092 W

250.018W

 

 

 

500°C

280.977 W

284.591W

 

 

 

600°C

313.708 W

317.994W

 

 

 

700°C

345.280 W

-

 

 

 

800°C

375.700 W

-

 

 

 

10

Image 12
Contents Omega Omega.comTM Safety Information IntroductionSpecifications ElectricalEnvironmental GeneralPower Switch F Selecting the Temperature Scale Main displayHold Mode only Main display Operating InstructionsMIN MAX with Time record Mode only Main display T1/ T2/T1-T2 Main display Input SelectionPt385/Pt3926 T1 Input RTD Probe Select Main display REL Relative mode only Main display F Selecting the Temperature Scale Second display 12 Hi/Lo Limits mode only Main display 10 T1/T2 /T1-T2 second display Input Selection11 Pt385/Pt3926 Input RTD Prob select second display Operator Maintenance Battery ReplacementRTD Probe Connection Temperature VS Resistance TABLEITS90 HH 504 / HH 505 Calibration Procedure Warranty / Disclaimer Temperature