SPECIFICATIONS
ELECTRICAL
Temperature Scale: Celsius or Fahrenheit
Measurement Range:
Pt385(100W )
Resolution: 0.1°C or 0.2°F
Accuracy: Accuracy is specified for operating temperatures over the range of 18°C to 28°C (64°F to 82°F), for 1 year, not including RTD probe error.
±(0.05% rdg + 0.2°C) on °C scale ±(0.05% rdg + 0.4°F) on °F scale
Temperature Coefficient: 0.1 times the applicable accuracy specification per °C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F).
Input Protection: 24V dc or 24V ac rms maximum input voltage on any combination of input pins.
Maximum Differential Common Mode Voltage (Maximum Voltage be- tween T1 and T2 during measurement): 1volt.
Reading Rate: one time per second.
Input Connector: MTP miniature RTD 3 prong connector.