Omega Engineering FSW-50 Electrical Wiring, Plain Washers Split Lock Washer Strain Relief NUT

Page 7

10398-6. GROMMET

10398-28. LEADSCREW

10398-30. LEADSCREW NUT

10398-29. LEADSCREW RETAINING SLEEVE

10398-27. ADJUSTING SPRING

10398-1. BODY

10398-36. 11/4" x 1" BUSHING SUPPLIED

10398-10. DIAPHRAGM CLAMP RING (SONICALLY WELDED)

3 RETAINING SCREWS

FLEXIBLE DRAG STRIP ASSEMBLY

IS AVAILABLE TO REPLACE DRAG

DISK AND DRAG DISK ARM. NO

FILTER BOOT IS USED.

10398-37. DRAG STRIP

38. CLAMP WASHER

39. DAMPENER WASHER

40. SCREW x 3/8" LONG

41. 2 PLAIN WASHERS

42. SPLIT LOCK WASHER

10398-5. STRAIN RELIEF NUT

10398-2. COVER

3.4 RETAINING SCREWS (NOT SHOWN)

10398-31. MICROSWITCH COMPLETE WITH 3 SLIP-ON TERMINALS

10398-24. MICROSWITCH BRACKET COMPLETE WITH SPRING LEVER

10398-32. 2 SCREWS

33.2 NUTS

34.2 SPLIT LOCKWASHERS

35.1 PLAIN WASHER

10398-4. O-RING

10398-25. 2 SCREWS

26. 2 SPLIT LOCKWASHERS

10398-22. SEAL SHAFT ASSEMBLY SUPPLIED ONLY COMPLETE WITH

A)SEAL SHAFT

B)DIAPHRAGM

C)DIAPHRAGM WASHER

D)FEED THRU SHAFT

10398-16. DRAG DISK ARM MODEL FSW-50

17.DRAG DISK ARM MODEL FSW-51

18.DRAG DISK ARM MODEL FSW-52

10398-19. DRAG DISK #1 (SMALL)

20.DRAG DISK #2 (MEDIUM)

21.DRAG DISK #3 (LARGE)

WARNING

DO NOT APPLY TORQUE TO DRAG DISK

ARM SINCE IT MAY BE UNTHREADED

FROM SEAL SHAFT (A).

Thus: 1. Teflon tape thread and tighten the plastic bushing into the pipe or tank fitting.

2.Teflon tape thread and tighten the switch into the PLASTIC bushing by applying a wrench to the hexagon section. Repeat steps 1 and 2 until the ARROW on the body points in the DIRECTION OF FLOW and threads are leak tight.

NOTE

Plumber’s tools such as pipe wrenches are not recommended. If possible, use a ‘Rigid’ type wrench where the smooth jaws closely fit the hexagon section.

Electrical Wiring

1.Remove gland nut, grommet and switch cover.

2.Strip outer jacket of electrical cord back approximately 11/4 inches. Strip insulation from individual conductors back approximately

1/4 inch.

3.Slip on terminals are supplied with each switch. Remove from switch terminals and crimp on or solder to electrical leads.

4.Feed electrical cable through gland nut, grommet and switch cover as shown.

5.Apply slip on terminals to appropriate contacts of microswitch.

6.Slide the cover down the cable and fasten to the body of the switch with 4 screws provided.

7.Slide the grommet down the cable until the outer jacket is level with the small end of the grommet.

8.Push the grommet into the tapered end of the cover. Hold the cable jacket to prevent rotation and thread the gland nut firmly on to the cover.

Figure 1: Wiring Schematic for power applied to the load when the flow level is less than set point (power to load interrupted when flow increases to above set point).

FSW-50 Series w/Standard Construction Parts Diagram

8

5

Image 7
Contents Data Acquisition TemperaturePRESSURE, Strain and Force FLOW/LEVELFor immediate technical or application assistance WARRANTY/DISCLAIMERReturn REQUESTS/INQUIRIES Servicing North AmericaGeneral Description SpecificationsFlow Range water calibrated @ 70˚F Drag Disk Size Available Drag DisksWorking Line Pressure Installation DimensionsAssembly Warning InstallationPlain Washers Split Lock Washer Strain Relief NUT Electrical WiringSwitch Point Adjustment