Omega Engineering HH505 manual HH 504 / HH 505 Calibration Procedure

Page 13

HH 504 / HH 505 CALIBRATION PROCEDURE

Note: The following calibration procedure should perform only by qualified technicians who have access to the items as following.

Equipment: The class of calibrator had better 10 times greater than the measured meter.

1.Turn off the meter then set the adjusted jumper to J1.

2.Insert the 383.00Ω calibrator to the T1 then turn on the meter, the main display reading reads 1830.0 then press " ENTER " key.

3.The second display is stabilized (about 5sec.), then press " ENTER " key when the meter auto power off then set the jumper back to J3 position.

4.Turn on the meter then Insert the 100.00Ω calibrator. The display reading reads 0°C if the calibration procedure is right.

G

+

_

383.00￿￿CALIBRATOR

 

+ RW1

 

 

_ RW2

383.00￿

 

 

 

G RW3

Pt385

Pt3926

-200

18.521Ω

16.996Ω

-100

60.256Ω

59.479Ω

0

100.000Ω

100.000Ω

100

138.505Ω

139.272Ω

200

175.856Ω

177.362Ω

300

212.052Ω

214.275Ω

400

247.092Ω

250.018Ω

500

280.977Ω

284.591Ω

600

313.708Ω

317.994Ω

700

345.280Ω

800

375.700Ω

RW1=RW2=RW3

RW=WIRERESISTANCE

1

Image 13
Contents Omega Omega.comTM Introduction Safety InformationElectrical SpecificationsGeneral EnvironmentalRelative value Recall display mode Power SwitchOperating Instructions Pt385/Pt3926 T1 Input RTD Probe Select MIN MAX with Time record ModeREL Relative mode Page Lo Limit value Recall display mode Hi Limit value Recall display mode12 Hi/Lo Limits mode only Main display Battery Replacement Operator MaintenanceRTD Probe Connection Temperature VS Resistance TABLEITS90 HH 504 / HH 505 Calibration Procedure Warranty / Disclaimer Temperature