Omega Engineering HH502 manual Calibration Procedure

Page 12

CALIBRATION PROCEDURE

Note: The following calibration procedures should be performed only by qualified technicians who have access to the following:

Equipment: The class of calibration should be 10 times greater than the measured meter.

Step 1: Short the J3 with jumper.

Step 2: Apply 65 mV to the T1 input from Volts Calibrator and wait till the reading becomes stable, then press the SET key until the display shows 0.1.

Step 3: Apply Type K 0°C to the input T2 and wait till the reading becomes stable, then press the SET key until the meter shuts off by itself.

Step 4: Remove the jumper from J3.

10

Image 12
Contents HH502 Servicing North America Safety Information IntroductionSpecifications ElectricalEnvironmental Power Switch F Selecting the Temperature ScaleHold Mode only Main display Operating InstructionsMIN MAX with Time record Mode only Main display T1 /T2/T1-T2 Main display Input SelectionJ T1 Input Thermocouple Type Select Main display REL Relative mode only Main display Sec. Minu. Selecting the Time scale 12 Hi/Lo Limits mode only Main display 10 T1/T2/T1-T2 second display Input Selection11 K/J T2 Input thermocouple type select second display Operator Maintenance Battery ReplacementCalibration Procedure Page Page WARRANTY/ Disclaimer Temperature