A&D UA-853 manual Specifications

Page 24

 

SPECIFICATIONS

 

 

Model

UA-853

Type

Oscillometric

Display

Digital, 20-mm character height

 

Pressure/pulse displayed simultaneously

Time/Date

12 hour format (AM/PM),

 

year (2006-2056), month, and day with

 

automatic adjustment for leap years

Memory

90 readings

Measurement range

Pressure: 20 mmHg to 280 mmHg

 

Pulse: 40 pulses to 180 pulses

Accuracy

Pressure: ±3 mmHg or ±2%,

 

whichever is greater

 

Pulse: ±5%

Pressurization

Automatic, using micropump

Depressurization

Active exhaust valve

Power source

4 type “AA” (1.5 volt) alkaline

 

batteries (not included) or 120 V

 

AC Adapter

Battery life

Approximately 4 months with

 

one daily measurement

Operating environment

50ºF to 104ºF (10ºC to 40ºC)

 

Less than 85% relative humidity

Storage environment

-14ºF to 140ºF (-10ºC to 60ºC)

 

Less than 95% relative humidity

Dimensions

Length: 4.9”(126 mm)

 

Width: 5.9”(150 mm)

 

Height: 6.1” (156 mm)

Weight

12.3 oz. (350 g) without batteries

Blood pressure measurements determined by the UA-853 are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation method within the limits prescribed by the American National Standards Institute for electronic or automated sphygmomanometers.

In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991

E-21

Image 24
Contents Blood Pressure Monitor Precautions Important INFORMATIONFast. Easy. Accurate Table of Contents ? What Display Symbols Mean Display Condition/Cause RecommendedAction Monitor Components Back? HOW UA-853 Works Before YOU StartAM/PM TimeWise Technology Cuff Size Indicator On The Cuff Selecting the Correct Cuff SizeAdjusting the Built-in Clock Taking Your Blood PressureSetting the Alarm Tips for Blood Pressure Monitoring Now you are ready. Follow these simple stepsImportant Measure blood pressure at the same time each day Measurement with the desired systolic pressure What is AN Irregular Heartbeat About Cuff Inflation MeterAbout Pressure Rating Indicator About Average ReadingAbout Memory Clearing the memory dataMemory + AM button Memory + PM buttonUsing the Monitor with AC Adapter Connecting the AC adapter to the monitorDisconnecting the AC adapter from the monitor What Is Blood Pressure? About Blood PressureWhat Affects Blood Pressure? What Causes Variations In Blood Pressure?Assessing High Blood Pressure for Adults JNC7 Classification Table for adults within the U.SWhat Is Hypertension? Why Measure Blood Pressure at Home?Can Hypertension Be Controlled? Keys to Successful Monitoring Establishing Baseline Measurements How Do I Record My Blood Pressure?Customer Support Tools Online Troubleshooting Problem Probable Cause Corrective ActionSpecifications Warranty. See warranty card for details Contact InformationWe Can Help