Omron Healthcare NE-C30 instruction manual Important Safety Notes, Operating the Device

Page 5

IMPORTANT SAFETY NOTES

OPERATING THE DEVICE (continued)

Do not insert any object into the compressor.

Make sure the air filter is clean. If the air filter has changed color or has not been used for 60 days, replace the filter.

Make sure the nebulizer kit is correctly assembled, the air filter is properly installed, and the air tube is correctly connected to the compressor and the nebulizer kit. Air may leak from the air tube during use if not securely connected.

Inspect the compressor (main unit) and the nebulizer parts each time before using the device. Make sure no parts are damaged, the nozzle and air tube are not blocked and the compressor operates normally.

Do not use the device if the air tube is bent.

Do not block the air filter cover.

Do not alter the baffle, the nozzle in the medication cup or any part of the nebulizer kit.

Do not add more than 7ml of medication to the medication cup.

Do not operate the device at temperatures greater than +104˚F (+40˚C).

Do not tilt the nebulizer kit so the angle of the kit is greater than 45˚. Medication may flow into the mouth.

Do not shake the nebulizer kit while using the device.

Do not subject the compressor, rechargeable battery, battery charger or any of the components to strong shocks, such as dropping on the floor.

This device is approved for human use only.

Do not disassemble or attempt to repair the device or components.

Operate the device only as intended. Do not use the device for any other purpose.

Dispose of the device, components and optional accessories according to applicable local regulations. Unlawful disposal may cause environmental pollution.

Use only Omron authorized parts and accessories. Parts and accessories not approved for use with the device may damage the unit.

Changes or modification not approved by Omron Healthcare will void the user warranty.

Make sure that the air tube is securely attached to the compressor (main unit) and

nebulizing parts, and does not come loose. Twist the air tube slightly when inserting it

into the connectors to avoid the tube disconnecting during use.

5

Image 5
Contents Compact Compressor Nebulizer Contents Staple your purchase receipt here IntroductionOperating the Device Safety InformationOperating the Device Important Safety NotesRisk of Electrical Shock Maintenance and Storage Main Unit Know Your UnitOptional Accessories Preparing the Nebulizer for USE Max Attach the mask to the inhalation top Attaching the AIR Tube Hold the nebulizer kit as illustrated on the right Using the DevicePress the power switch again to stop nebulization Power Indicator Cleaning After Each USE Daily Disinfecting Medical Disinfectant Method Insert a new air filter into the air filter cover Changing the AIR FilterCaring for the Device Storing the Device Problem Cause Solution Troubleshooting GuideLimited Warranties WarrantyFCC Statement Contents SpecificationsPage Omron Healthcare, Inc