UNIT DISASSEMBLY & MAINTENANCE
NOTE: The mesh cap may be damaged by hardened medication that is left in the vibration horn effecting how the unit nebulizes. The following are maintenance recommendations after each use to keep your unit working at its best level.
1. | NOTE: Be sure that the unit is turned OFF. |
| Remove the medication cup and dispose of any |
| remaining medications. |
2. | Be sure to nebulize all medication from the unit. |
| The medication cup needs to be removed while |
the unit is on to remove all medication that might be in the mesh cap or transducer. The unit will no longer provide any aerosol when these parts are completely empty.
NOTE: The lack of any misting will occur with a high pitched sound when all medication has been nebulized. Quickly turn the unit off when this sound is heard it shows that there is no more medication in the unit.
3. | Pour a mixture of one part vinegar and three |
| parts distilled water into the medication cup |
| after each use. Nebulize the mixture for about |
45 to 60 seconds. Remove the medication cup and run the unit until it can no longer mist. Turn the unit off at this point.
NOTE: Make sure that there is no medication left on the mesh cap or horn that can dry and cause damage or breakage to the mesh cap when removed from the unit.
Place
We recommend that a small bottle of the water/vinegar mixture be kept with your unit at all times for proper cleaning.