Omron NE-U03V manual Unit Disassembly & Maintenance

Page 9

UNIT DISASSEMBLY & MAINTENANCE

NOTE: The mesh cap may be damaged by hardened medication that is left in the vibration horn effecting how the unit nebulizes. The following are maintenance recommendations after each use to keep your unit working at its best level.

1.

NOTE: Be sure that the unit is turned OFF.

 

Remove the medication cup and dispose of any

 

remaining medications.

2.

Be sure to nebulize all medication from the unit.

 

The medication cup needs to be removed while

the unit is on to remove all medication that might be in the mesh cap or transducer. The unit will no longer provide any aerosol when these parts are completely empty.

NOTE: The lack of any misting will occur with a high pitched sound when all medication has been nebulized. Quickly turn the unit off when this sound is heard it shows that there is no more medication in the unit.

3.

Pour a mixture of one part vinegar and three

 

parts distilled water into the medication cup

 

after each use. Nebulize the mixture for about

45 to 60 seconds. Remove the medication cup and run the unit until it can no longer mist. Turn the unit off at this point.

NOTE: Make sure that there is no medication left on the mesh cap or horn that can dry and cause damage or breakage to the mesh cap when removed from the unit.

Place 2-3 drops of the water/vinegar mixture onto the mesh cap surface when dried medica- tion is noted. Nebulize distilled water from the medication cup for at least 30 seconds.

We recommend that a small bottle of the water/vinegar mixture be kept with your unit at all times for proper cleaning.

9

Image 9
Contents Portable Nebulizer Model NE-U03V Introduction Table of ContentsRead ALL Instructions Before Using Important SafeguardsKnow Your Unit Page Page How to Assemble Unit AssemblyOperating Instructions Unit Disassembly & Maintenance Nebulizer Parts and Casing CleaningAfter Every Cleaning Mesh CleaningMesh Cap StorageUsing Batteries Battery OperationCheck before sending for service Trouble Shooting GuideAvailable Accessories Specifications Common Questions about the Microair How often should I clean the MicroAir?Will the mesh break easily? Licensed Under U.S. PAT TWO Year Limited WarrantyOmron HEALTHCARE, INC