Tanita 4521 specifications Troubleshooting, Transporting and Storing

Page 7

TROUBLESHOOTING

The following guide will help you to solve common problems:

PROBLEM

SOLUTION

 

 

Incorrect patient

A. Platform is too low and hitting platform feet or

weight readings

carpet. Increase platform height. See “Leveling

 

the Platform” on page 2.

 

B. Person, other than person being weighed,

 

is touching or leaning on scale handlebar

 

C. Weight of wheelchair or chair is not taken into

 

account. See “Operating Instructions” on page 4.

 

D. Platform is not level. See “Leveling the Platform”

 

on page 2.

 

 

Patient weight readings

Check scale calibration and calibrate if necessary.

consistently too high

See “Calibration” on page 5.

or too low

 

 

 

Display flashes “Lo-b”

Recharge battery. See “Charging the Battery” on

 

page 3.

 

 

TRANSPORTING AND STORING

See Fig. 5 for visual assistance.

1.

Unplug the adapter

Fig. 5

 

from the outlet, then

 

 

unscrew it from the

 

 

electronic control

 

 

box and set the

 

 

adapter aside.

 

2.

Slide the adjustable

 

 

rail to its widest setting.

 

3.

Loosen the locking

 

 

and slide it off the

.

steppaipanpngedPoawunrsasshibctea“uflOnoetNriel”zbeurotton

ZERO

POUNDS

KILOS

4.Slowly lift the scale by the handle. Avoid letting the rails bang together.

5.Wrap an elastic cord (64010) around each end of the rails and hook the cord ends together.

6.Unscrew the scale cable from the control box and firmly place the control panel box on the velcro tape of the center column. Wrap the cable around the center column above the control panel box and secure.

7.The scale is now ready for transportation or storage.

8.Remember to keep the adapter with the scale.

9.For additional convenience and protection of your portable wheelchair scale, a scale storage bag is available from your Tanita representative.

6

Image 7
Contents As a Wheelchair Scale As a Chair Scale Charging the BatteryControl Panel Box Special Features Specifications Cleaning CalibrationDo not leave patients unattended on the scale Leveling the Platform Assembly InstructionsBefore Assembly Assembling the Portable Wheelchair ScaleProduct Description Battery ChargingCharging the Battery Special FeaturesAs a Chair Scale SpecificationsOperating Instructions As a Wheelchair ScaleCleaning Care and MaintenanceCalibration Following guide will help you to solve common problems TroubleshootingTransporting and Storing Computer Board Parts ListHow to Obtain Warranty Service WarrantyTanita Corporation of America, Inc