Problem/Possible cause

Solution

 

Displays error message: High leak, please check system setup and all connections

There is excessive leak.

Adjust position of mask and headgear.

 

Air tubing is not connected properly. Connect firmly at both

 

ends.

The following message is displayed on the LCD after you try to update settings or copy data to the SD card: Card error, please remove SD card and contact service provider

SD card is not inserted correctly.

Ensure that the SD card is inserted correctly.

You may have removed the SD card

Reinsert the SD card and wait for the home screen or the

before settings were copied to the

Settings updated successfully, press any key message

CPAP device.

to appear on the LCD.

 

Note: This message only appears once. If you re-insert the

 

SD card after you have updated your settings, the message

 

will not be re-displayed.

The following message is NOT displayed on the LCD after you try to update the settings using the SD card: Settings updated successfully, press any key

The settings were not updated.

Contact your clinician/service provider immediately.

SD card

An SD card has been supplied to gather therapy data from your S9 device and provide settings updates from your clinician. When instructed to do so, disconnect the S9 device from the power outlet, remove your SD card, insert it in the protective folder and send it to your clinician. For more information on removing and inserting your card refer to the S9 SD Card Protective Folder provided with your device. Please retain the S9 SD Card Protective Folder for future use.

Technical specifications

Operating pressure range

4 to 20 cm H2O

Maximum single fault steady state

30 cm H2O

pressure

 

Pressure measurement tolerance

±0.5 cm H2O ± 4% of the measured reading

Flow measurement tolerance

±6 L/min or 10% of reading, whichever is greater

5 English

Image 6
Contents S9 Escape Contraindications Masks and humidifiers SetupS9 Escape indications for use Adverse effectsNavigating the menus Cleaning and maintenanceControl panel Getting startedTroubleshooting Problem/Possible causeProblem/Possible cause Solution Technical specifications SD cardPage Environments Servicing SymbolsPneumatic flow path Environmental informationLimited warranty English English