Problem/Possible cause | Solution |
| |
Displays error message: High leak, please check system setup and all connections | |
There is excessive leak. | Adjust position of mask and headgear. |
| Air tubing is not connected properly. Connect firmly at both |
| ends. |
The following message is displayed on the LCD after you try to update settings or copy data to the SD card: Card error, please remove SD card and contact service provider
SD card is not inserted correctly. | Ensure that the SD card is inserted correctly. |
You may have removed the SD card | Reinsert the SD card and wait for the home screen or the |
before settings were copied to the | Settings updated successfully, press any key message |
CPAP device. | to appear on the LCD. |
| Note: This message only appears once. If you |
| SD card after you have updated your settings, the message |
| will not be |
The following message is NOT displayed on the LCD after you try to update the settings using the SD card: Settings updated successfully, press any key
The settings were not updated. | Contact your clinician/service provider immediately. |
SD card
An SD card has been supplied to gather therapy data from your S9 device and provide settings updates from your clinician. When instructed to do so, disconnect the S9 device from the power outlet, remove your SD card, insert it in the protective folder and send it to your clinician. For more information on removing and inserting your card refer to the S9 SD Card Protective Folder provided with your device. Please retain the S9 SD Card Protective Folder for future use.
Technical specifications
Operating pressure range | 4 to 20 cm H2O |
Maximum single fault steady state | 30 cm H2O |
pressure |
|
Pressure measurement tolerance | ±0.5 cm H2O ± 4% of the measured reading |
Flow measurement tolerance | ±6 L/min or 10% of reading, whichever is greater |
5 English