Invacare HF2POST9, HF2POST6, HF2POSTD, 1148114 manual Filling the Cylinder

Page 13

SECTION 4—INSPECTION AND COMPRESSOR INTERFACE

2.Ensure toggle valve is open. Open the toggle valve by rotating counterclockwise.

3.Examine the gauge on the connected conserver or regulator. Perform one of the following:

If the cylinder pressure is less than 1500 psig, fill the cylinder. Connect the cylinder to the compressor. Refer to Filling the Cylinder on page 13.

If the cylinder pressure is greater than 1500 psig., DO NOT attempt to fill this cylinder. It may not fill. Use cylinder as normal.

Filling the Cylinder

NOTE: For this procedure, refer to FIGURE 4.2 thru FIGURE 4.7.

1.Check the cylinder pressure before attempting to fill cylinder. Refer to Checking Cylinder Pressure on page 12.

2.Close the toggle valve by rotating clockwise (FIGURE 4.2).

3.Remove the conserver or regulator according to the device manufacturer instructions (FIGURE 4.2).

Part No 1148114

13

HomeFill

Image 13
Contents HomeFill Oxygen Post Valve Cylinder HomeFill Table of Contents Receiving Inspection Special NotesIntended USE General Guidelines General Guidelines General Guidelines Product Diagram FeaturesTypical Product Parameters DimensionsCylinder Prefill Inspection Inspection and Compressor InterfaceExternal Examination Checking Cylinder Pressure Filling the Cylinder Component Identification Resetting Connector Fillport Connecting the Cylinder to the Compressor Disconnecting the Cylinder from the Compressor Connector Fillport Maintenance Cylinder Fill Times TroubleshootingHomeFill Limited Warranty Limited Warranty USA