SECTION 4—INSPECTION AND COMPRESSOR INTERFACE

CONSERVER/REGULATOR

INSTALLED

Gauge

Toggle

 

Valve

Conserver/

Regulator

Cylinder

CONSERVER/REGULATOR

REMOVED

FIGURE 4.2 Component Identification

4.Remove the post valve cover (located on the cylinder) and the connector fillport covers (located on the compressor) (FIGURE 4.3).

Post Valve Cover

Connector

Fillport Cover

FIGURE 4.3 Fillport Covers

5.On the compressor unit, momentarily push DOWN on the outer ring (sleeve) of the connector fillport to reset the connector (FIGURE 4.4).

HomeFill

14

Part No 1148114

Image 14
Contents HomeFill Oxygen Post Valve Cylinder HomeFill Table of Contents Intended USE Special NotesReceiving Inspection General Guidelines General Guidelines General Guidelines Features Product DiagramDimensions Typical Product ParametersInspection and Compressor Interface Cylinder Prefill InspectionExternal Examination Checking Cylinder Pressure Filling the Cylinder Component Identification Resetting Connector Fillport Connecting the Cylinder to the Compressor Disconnecting the Cylinder from the Compressor Connector Fillport Maintenance Troubleshooting Cylinder Fill TimesHomeFill Limited Warranty Limited Warranty USA