DISPLAY MESSAGES

The thermometer has been designed to display feedback messages to ensure proper usage and accurate measurement. In the event

of a malfunction or incorrect measurement, the following messages will be displayed:

Displayed

Cause

 

 

The measured temperature is below 95°F (35°C)

 

or above 107.6°F (42°C).

 

Instructions

 

1. Check correlation between the selected mode icon

 

and the actual measurement location used.

 

2. Turn the unit OFF and then ON again.

 

3. Retake temperature using the proper technique

 

for oral, axillary, and rectal method.

 

4. Any of the methods may be used for accuracy

 

depending on individual circumstances.

 

 

Displayed

Cause 1

 

The thermometer is too cold or too hot to operate.

 

Operating temperature range is 50-104°F (10-40°C)

 

Cause 2

 

The thermometer was placed in the measurement

 

method before the “hourglass” symbol disappeared

 

and before two “Beeps” were heard.

Instructions

1.Bring the thermometer to a room where the temperature is between 50°F and 104°F (10°C and 40°C). Thermometer should be allowed to stabilize at room temperature for at least 45 minutes.

2.Wait at least one minute and then turn the thermometer ON again.

3.Wait for the “hourglass” symbol to disappear and two “beeps”

to be heard.

4. Retake temperature using the proper technique. (as described on page 9)

11

Image 11
Contents WAY Instant Thermometer Table of Contents Thermometer is not Water Resistant How does it work? IntroductionComponents Know Your Thermometer To install or replace the battery Battery Installation ReplacementOperation Instructions General InstructionsRectal Method Oral MethodAxillary underarm Method Operation in a Warm Environment Changing Scale Between C and FDisplayed Display MessagesUnit malfunction Care and Cleaning Accuracy Product SpecificationsLimited ONE Year Warranty Distributed by