Omron Healthcare MC-600 manual Unit malfunction

Page 12

DISPLAY MESSAGES

Cause 1

The probe was inappropriately placed as indicated by measurement method, example: oral, axillary, and rectally.

Cause 2

Temperature did not register because of poor contact or placement.

Instructions

1.Check correlation between selected mode icon and actual measurement location used.

2.Turn the unit OFF, wait one minute, Turn the unit on and retake the temperature using the proper technique. (see page 9).

Cause

Unit malfunction.

Instructions

Stop using the thermometer – contact Omron

Customer Service 1-800-634-4350

Cause

Temperature was not taken within 1 minute after the thermometer reached the ready state.

Instructions

Turn the unit OFF. Turn it ON again after one minute to retake the temperature (as described on page 9).

Cause

Low battery.

Instructions

The battery has enough remaining life

for approximately 10 more measurements.

Cause

The battery is very low

Instructions

No measurement can be taken.

Replace the batter with a new one.

12

Image 12
Contents WAY Instant Thermometer Table of Contents Thermometer is not Water Resistant Introduction How does it work?Components Know Your Thermometer Battery Installation Replacement To install or replace the batteryGeneral Instructions Operation InstructionsOral Method Axillary underarm MethodRectal Method Changing Scale Between C and F Operation in a Warm EnvironmentDisplay Messages DisplayedUnit malfunction Care and Cleaning Product Specifications AccuracyLimited ONE Year Warranty Distributed by