TROUBLESHOOTING/ERROR CODES
Condition | Possible Cause | Corrective Action |
|
|
|
Consistent Low | Temperature reading | The thermometer measurement |
temperature readings/ | is below | range is 32° – 210.2°F (0°C – 99°). |
Displays “Lo” | measurement range | Contact your physician or if you |
| (below 32°F or 0°C). | have or suspect any medical |
|
| problems. |
| The lens may be | Clean the lens gently with a cotton |
| dirty. | |
| swab with a small amount of | |
|
| |
| The thermometer | alcohol on it. |
| Repeat temperature measurement | |
| may not be | |
| positioned properly. | procedures. |
|
|
|
Consistent high | Temperature reading | The thermometer measurement |
temperature readings/ | is above the | range is 32° – 210.2°F (0° – 99°C) |
Displays “Hi” | measurement range | Contact your physician if you have or |
| (above 210.2°F or | suspect any medical problems. |
| 99°C). |
|
|
|
|
Consistent error | Operating | The product operating range |
messages/Displays “Err” | temperature is not | temperature is 60°– 95°F |
| within the permissible | (16°– 35°C). Move to an area within |
| range for | the above specifications, allow the |
| thermometer to adjust to the | |
| measurement. | |
| environment and repeat temperature | |
|
| |
|
| measurement procedures. |
|
|
|
Low battery warning | Battery power | Replace battery. See the“Inserting |
| is low. | and Replacing Batteries” section of |
|
| this manual. |
|
|
|
Low battery warning | Battery power is | Replace battery immediately. See |
with “Lo” symbol | too low for the | “Inserting and Replacing Batteries” |
| thermometer to | section of this manual. |
| function. |
|
|
|
|
PRODUCT SPECIFICATIONS
Model Number | Model Number | |
Power Source | 2 AAA batteries | |
|
| |
Measurement Method | Infrared | |
|
| |
Measurement Range | 32° – 210.2°F (0° – 99°C) | |
|
| |
| Person Mode: ± 0.4°F (±0.2°C) between 98.6°F – | |
Accuracy | 102.2°F (36°C – 39°C); and ± 0.5°F (±0.3°C) at less | |
than 98.6°F (36°C) or greater than 102.2°F (39°C). | ||
| ||
| Object Mode: ± 5% | |
|
| |
Memory Capacity | 20 memories, 10 in person mode, 10 in object mode | |
|
| |
Sleep Mode | 1 minute after last button operation | |
|
| |
Operation Environment | Temperature: 60.8° – 104°F (16° – 14°C) Humidity | |
up to 95% | ||
| ||
|
| |
Storage Environment | Temperature: | |
95% | ||
| ||
|
|
Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no circumstance should you disassemble or attempt to repair the unit by yourself. Contact information for HoMedics Consumer Relations Department can be found on the warranty page. These specifications are subject to change without notice.
21 | 22 |