Omega Vehicle Security HH507 manual HH 506R/HH 507R Calibration Procedure

Page 13

HH 506(R)/HH 507(R) CALIBRATION PROCEDURE

Note: The following calibration procedure should perform only by qualified technicians who have access to the items as following.

Equipment: The class of calibrator had better 10 times greater than the measured meter.

1.Set the jumper position on J1 position.

2.Input DCV 17mV to the T1 after the display is stabilized (forward 4 digits), then press “ HOLD “ key and "0.0" is displayed in main display.

3.Input DCV 60mV after the display is stabilized (first 4 digits), then press

HOLD “ key and "0.1" is displayed in main display.

4.Input 0￿C (K-type) after the display is stabilized (first 4 digits), then press

HOLD “ key and "0.2" is displayed in main display.

5.Don't push any key and the meter will turn off automatically after 20 seconds.

6.After adjustment set the jumper back to J3 position.

7.Turn on the meter then input 0￿C (K-type). The display reading reads 0°C if the calibration procedure is right.

1

Image 13
Contents Omega Omega.comTM Introduction Safety InformationElectrical SpecificationsT/E-TYPE Environmental F Selecting the Temperature Scale Power SwitchHold Mode only Main display Operating InstructionsMIN MAX with Time record Mode only Main display Timer STOP-WATCH ModeJ/T/E/R/S/N T1 Input Thermocouple Type Select Main display REL Relative mode only Main display Hi/Lo Limits mode only Main display 11 Hi/Lo Limits value Recall display mode Relative value Recall display modeButton Battery Replacement Operator MaintenanceHH 506R/HH 507R Calibration Procedure Warranty / Disclaimer 2813/0799