Citizen CT810 instruction manual Battery Replacement, After USE and Storage

Page 6

ENGLISH

BATTERY REPLACEMENT

1.Replace with new battery if the battery indicator ( ) lights up when the switch is pressed.

2.Use a Phillips screwdriver to remove the screw.

3.Use a coin to rotate the battery cover in the direction of the arrow, then remove the cover. (Fig. 9)

4.Remove the old battery, and replace it with a new one. (Fig. 10)

5.Reinstall the battery cover, and retighten the screw. (Fig. 11)

Fig. 9

Fig. 10

Fig. 11

(Be sure to orient the battery correctly)

Caution

1.Remove then reinstall the battery if an error message is displayed, the display is abnormal, or nothing is displayed when the switch is pressed after battery replacement.

2.To protect the environment, dispose of the used battery according to local regulations.

3.Remove the battery if the ear thermometer will not be used for a long time.

AFTER USE AND STORAGE

-After use, clean the probe and probe window with cotton soaked in alcohol or wet tissue available on the market, then wipe with a dry cloth.

-Storing the ear thermometer at 5 to 35 ˚C allows you to measure temperature at any time. Storing the ear thermometer below or above this temperature range above may cause it to display Error 4. In such a case, leave the ear thermometer at 5 to 35 ˚C for 10 minutes before measurement.

6

Image 6
Contents Digital EAR Thermometer Precautions About Ear Thermometers HOW to Operate / Memory Function If the ambient temperature is 35 ˚C or higher Auto Power OFF FunctionBattery Replacement After USE and StorageCleaning Description of Display MarkQuestions and Answers Q3 Can I measure temperature without the probe cover?Specifications Accuracy