Citizen CT810 instruction manual Cleaning, Description of Display Mark

Page 7

-Do not store the ear thermometer where it will be exposed to moisture, direct sunlight, high temperature or high humidity, or excessive dust.

CLEANING

Do NOT store the ear thermometer soiled or moist. If the ear thermometer is soiled, take the following steps to clean it.

1. Use a soft sponge soaked in neutral detergent to lightly wipe dirt off the probe and probe window.

2. Rinse the ear thermometer in standing lukewarm water for less than 20 seconds to remove dirt and detergent.

3. Be sure to carefully dry the ear thermometer with a soft cloth.

- Avoid using the ear thermometer for measurement immediately after washing. Wait at least 5 minutes before use.

- Do NOT rinse the ear thermometer directly under strong water flow from a faucet.

- Do NOT use a toothbrush, scouring sponge or pad, steel wool, cleanser, etc. to clean this instrument.

DESCRIPTION OF DISPLAY MARK

Error 1 This message will be displayed if the thermometer fails to make stable measurement of body temperature within 30 seconds. Insert the probe into the ear promptly after pressing the switch.

Error 2 This message will be displayed if the switch is pressed with the thermometer in the ear. Be sure to press the switch and confirm that “Fig. 7 ( )” is displayed, then insert the probe into the ear for measurement.

Error 3 This message will be displayed if the temperature measured is above 42 ˚C. The thermometer may have measured the temperature of a nearby heat source. Avoid external heat sources and try again.

Error 4 This message will be displayed if the ambient temperature is out of the measurement range (below 5 ˚C or above 40 ˚C). Move to a place within the proper temperature range (from 5 to 40 ˚C) end let the thermometer stand for at least 10 minutes before measuring again.

ENGLISH

7

Image 7
Contents Digital EAR Thermometer Precautions About Ear Thermometers HOW to Operate / Memory Function Auto Power OFF Function If the ambient temperature is 35 ˚C or higherAfter USE and Storage Battery ReplacementDescription of Display Mark CleaningQ3 Can I measure temperature without the probe cover? Questions and AnswersAccuracy Specifications