TROUBLESHOOTING

LCD screen

Description & Solution

fl ashes

Battery power is low, but the CTD504 can still measure.

Remove the worn-out battery. Insert a new CR2032 battery as soon as possible. Without enough energy, the CTD504 cannot function as it should.

lights

Battery power is too low to measure.

Remove the worn-out battery. Insert a new CR2032 battery as soon as possible.

Too high ambient temperature

Ambient temperature is higher than 40.0 °C (104.0 °F)

Leave the CTD504 in a room around 10.0 °C – 40.0 °C (50.0 °F – 104.0 °F) for 30 minutes. Then measure again.

Too low ambient temperature

Ambient temperature is lower than 10.0 °C (50.0 °F)

Leave the CTD504 in a room around 10.0 °C – 40.0 °C (50.0 °F – 104.0 °F) for 30 minutes. Then measure again.

LCD screen

Description & Solution

 

Outside displayed temperature range I

Result is higher than 43.0 °C (109.4 °F).

 

English

 

 

The user should rest in a steady room temperature for at

 

least 30 minutes. Then measure again.

 

Español

Outside displayed temperature range II

 

Result is lower than 34.0 °C (93.2 °F).

 

 

 

The user should rest in a steady room temperature for at

Português

least 30 minutes. Then measure again.

 

 

 

Error symbol I

 

 

Disturbance occurred while measuring.

 

Deutsch

Error symbol II

 

Repeat the measuring steps again. If the error persists,

 

return the CTD504 to your local distributor.

 

 

Repeat the measuring steps again. If the error persists,

Italiano

Memory error occurred while measuring.

 

 

return the CTD504 to your local distributor.

 

Français

OTHER INFORMATION

 

 

 

This product conforms to the provisions of the EC directive

 

 

93/42/EEC (Medical Device Directive). And also complies

0197

 

with following standards (included but not limited):

中文

 

Safety standard:

 

 

 

EN 60601-1 Medical electrical equipment part 1: General requirements for safety

 

EMC standard:

 

Русский

EN 60601-1-2 Medical electrical equipment part 1-2: General requirements for

 

safety- Collateral standard: Electromagnetic compatibility- Requirements and tests

Performance standards:

EN 12470-5, Clinical thermometers - Part 5: Performance of infra-red ear thermometers (with maximum device)

Follow instructions for use.

+40°C

 

(104°F)

The upper and lower limits of temperature.

+10°C

 

(50.0°F)

 

Keep dry.

– Eng 6 –

– Eng 7 –

Image 5
Contents For Digital EAR Thermometer Description Precautions Intended UseHOW to Switch from to from to HOW to OperateBattery Replacement Recall MemoryMaintenance and Storage Cleaning InstructionsDisinfection Probe & Exterior Area Fever AlertOther Information TroubleshootingRecalibration Notice Warranty For One Year from Manufacturing DateSpecifications