Safety Regulations (cont'd.)
Light sources: safety regulations Protective measures of the manufacturer:
★★ UV protection screen in front of the speci- men plane prevents the user from looking directly into the UV rays.
★★ Dummy filter carriers in the free positions of the rapid filter changer prevent direct UV radiation from reaching the eyes.
★★ UV filters are installed in the observation beam paths to protect the eyes.
★★ The
Warning
★★ UV radiation could damage the eyes. Therefore:
★★ Never look into the light spot on the speci- men plane without a UV protection screen.
★★ Never look into the eyepieces if no excita- tion filter is in the beam path.
★★ Fill empty filter positions with dummy filter carriers (e.g. M165 FC).
★★ Do not select a white, strongly reflective background for the specimen.
Supply unit
Always unplug the supply unit from the power supply:
★★ When installing and disassembling the lamp housing
★★ Before opening the lamp housing
★★ When replacing the
★★ During maintenance work on the supply unit
Leica | User Manual | 7 |
|
|
|