Omega Vehicle Security LDP1XX-21 manual Adjustment and Calibration Procedure, RTD Models

Page 11

11.- ADJUSTMENT AND CALIBRATION PROCEDURE

11.1.- RTD MODELS

A decade box or RTD simulator is required.

1.- Check that the instrument is not powered.

2.- Connect the RTD simulator or decade box to the signal input connector (See paragraph 8.3).

3.- Remove the rear cover taking off all the screws to obtain access to the control board.

(See page 7 paragraph 7).

4.- Power up the instrument with the appropriate power supply and wait until 3 minutes before to start the adjust.

5.- Adjust the RTD simulator to generate the low signal indicated in TABLE-A.

6.- Display must shows the DISPLAY value indicated in TABLE-A, if not turn the trimmer PT1 located on the Signal Conditioner (See paragraph 10.1) until display shows the DISPLAY value.

7.- Adjust the RTD simulator to generate the high signal, see TABLE-B.

8.- Display must shows the DISPLAY value indicated in TABLE-B, if not turn the trimmer PT2 located on the Signal Conditioner (See paragraph 10.1) until display shows the DISPLAY value.

9.- Repeat steps 5 to 8 until the two values are correct.

10.- Check that the linearity remains correct, testing some other points of the range.

11.2.- THERMOCOUPLE MODELS

A Thermocouple simulator with Cold Junction compensation is required.

1.- Check that the instrument is not powered.

2.- Connect the Thermocouple simulator to the signal input connector (See paragraph 8.4).

3.- Remove the rear cover taking off all the screws to obtain access to the control board. (See page 7 paragraph 7).

4.- Power up the instrument with the appropriate power supply and wait until 3 minutes before to

start the adjust.

5.- Adjust the Thermocouple simulator to generate the low signal indicated in TABLE-A.

6.- Display must shows the DISPLAY value indicated in TABLE-A, if not turn the trimmer PT1located on the Signal Conditioner (See paragraph 10.2) until display shows te correct value.

7.- Adjust the Thermocouple simulator to generate the high signal indicated in TABLE-B.

8.- Display must shows the DISPLAY value, if not turn the trimmer PT2 located on the Signal Conditioner (See paragraph 10.2) until obtain the DISPLAY value.

9.- Repeat steps 5 to 8 until the low and the high signals are correct.

10.- Check that the linearity remains correct, testing some other points of the range.

TABLE - A

MODEL

LOW SIGNAL

DISPLAY

 

 

 

 

 

 

M1

0°C

000.0

M2

0°F

0000

M3

0°F

000.0

J1

0°C

0000

J2

32°F

0032

K1

0°C

0000

K2

32°F

0032

T1

0°C

0000

T2

0°F

0000

E1

0°C

0000

E2

32°F

0032

R1

300°C

0300

R2

32°F

0032

S1

200°C

0200

S2

32°F

0032

L1

0°C

0000

 

 

 

TABLE - B

MODEL

HIGH SIGNAL

DISPLAY

 

 

 

 

 

 

M1

600°C

600.0

M2

500°F

0500

M3

200°F

200.0

J1

600°C

0600

J2

1000°F

1000

K1

1200°C

1200

K2

2000°F

2000

T1

400°C

0400

T2

400°F

0400

E1

600°C

0600

E2

1200°F

1200

R1

1700°C

1700

R2

3100°F

3100

S1

1700°C

1700

S2

3100°F

3100

L1

600°C

0600

 

 

 

Page : 10

Image 11
Contents Users Guide LDP-124 and LDP-144 Series For TemperatureUSA Contents Important Safety Considerations Reference KEY ENGINEERING´S customer service departmentUnpacking and Inspection LDPSensor Break Detection Main FeaturesMechanical ConnectionsLDP-1X4-MX ModelsEnvironmental General SpecificationsDisplay ANALOG-TO-DIGITAL ConversionSensor Connections for Model MX WiringPower SUPPLY, Recommended Wiring Protection FusesSignal Conditioners for Thermocouple Sensors Temperature RangesSignal Conditioners Signal Conditioner for RTD SensorThermocouple Models Adjustment and Calibration ProcedureRTD Models Mechanical Dimensions mm inches InstallationReturn REQUESTS/INQUIRIES WARRANTY/DISCLAIMERData Acquisition TemperaturePRESSURE, Strain and Force FLOW/LEVEL