Emerson Process Management instruction manual START-UP and Calibration, Model 381 ORP

Page 18

MODEL 381pH/ORP

SECTION 2.0

 

INSTALLATION

2.7START-UP AND CALIBRATION

2.8MODEL 381 pH and pHE

2.8.1SENSOR PREPARATION. Shake down the sensor to remove any air bubbles that may be present at the tip of the pH glass bulb. In most cases, the pH sensor can simply be installed as shipped and read- ings with an accuracy of ± 0.6 pH may be obtained. To obtain greater accuracy or to verify proper operation, the sensor must be calibrated as a loop with its com- patible analyzer or transmitter.

WARNING

Glass electrode must be wetted at all times (in storage and in line) to maximize sensor life.

2.8.2CALIBRATION USING BUFFER SOLUTIONS OR GRAB SAMPLES. The pH sensor-analyzer/ trans- mitter loop may be calibrated by submersing the sensor in a buffer solution (standard solutions of known pH val- ues) or in a process grab sample whose pH value maybe checked by a calibrated laboratory or portable pH meter. Please refer to the analyzer's or the trans- mitter’s respective instruction manuals for proper calibration procedures.

2.9 MODEL 381 ORP.

2.9.1MODEL 396P ORP. Most industrial applications have a number of ORP reactions occurring in sequence or simultaneously. There can be several components that are oxidized or reduced by the reagents that are used. Theoretically, the ORP poten- tial is absolute because it is the result of the oxidation- reduction equilibrium. However, the actual measured potential is dependent on many factors, including the condition of the surface of the ORP platinum elec- trode. Therefore, the sensor should be allowed 1-2 hours to become “conditioned” to the stream when first set-up or after being cleaned.

2.9.2 Calibration

CAUTION

The solution used during the following check is an acid and should be handled with care. Follow the directions of the acid manufactur- er. Wear the proper protective equipment. Do not let the solution come in contact with skin or clothing. If contact with skin is made, immediately rinse with clean water.

1.Make a temporary electrical connection between the sensor and the instrument.

2.Obtain a standard solution of saturated quinhydrone (PN R508-160Z). This can also be made quite sim- ply by adding a few crystals of quinhydrone to either pH 4 or pH 7 buffer. Quinhydrone is only slightly sol- uble therefore a few crystals will be required. (Refer to Section 4.3. for an alternate ORP standard solu- tion).

3.Immerse the sensor in the standard solution. Allow 1-2 minutes for the ORP sensor to stabilize.

4.Adjust the standardize control of the instrument to the solution value shown in Table 2-1. The resulting potentials, measured with a clean platinum electrode and saturated KCl/AgCl reference electrode, should be within ±20 millivolts of the value shown in Table 2- 1. Solution temperature must be noted to insure accurate interpretation of results. The ORP value of saturated quinhydrone solution is not stable over long periods of time. Therefore, these standards should be made up fresh each time they are used.

TABLE 2-1.

ORP of Saturated Quinhydrone Solution

(In Millivolts)

 

pH 4 Solution

pH 7 Solution

 

 

 

 

 

 

 

Temp °C

20

25

30

20

25

30

 

 

 

 

 

 

 

Millivolt Potential

268

264

260

94

87

80

 

 

 

 

 

 

 

5.Remove the sensor from the buffer, rinse and install in the process.

14

Image 18
Contents PH/ORP Sensor Essential Instructions Model 381 pH/ORP Sensor About This Document Section Description and Specifications Features and ApplicationsMeasuring Range pH Sensor SpecificationsMaterials of Construction Pressure Rating 790 kPa abs at 100C 100 psig at 212FExample This page left blank intentionally Section Installation Model 381 Sensor Component Locator Diagram Section Submersion Installation Diagram Flow Through Installation Diagram Code 03 or Retro-Fit Kit Code 04 Flow Powered Cleaner, Model 40038105 40038103 40038116 START-UP and Calibration Model 381 ORPCalibration Section Maintenance Temperature C Millivolt Potential 268 264 260 Troubleshooting Probable Cause Remedy Temperature vs Resistance of Auto TemperatureElement Accessories Model 381 Spare PartsThis page left blank intentionally Section Ultrasonic Cleaner Operation 40038156 40037505 NON Warranty Repair Section Return of MaterialReturn of Materials Request Warranty Return of MaterialSpecifications subject to change without notice