Emerson Process Management TF396 instruction manual LOW Flow Cell PN, Jet Spray Cleaner PN

Page 11

MODEL TF396

SECTION 2.0

 

INSTALLATION

LOW FLOW CELL PN 24091-00

Inlet and outlet connections are stainless steel and take 1/4-inch OD tubing. Flow cell is polycarbonate with 1/4-inch FNPT fittings.

WETTED MATERIALS:

BODY - ACRYLIC

NUT - CPVC

FITTINGS - 316 SST

SEALS - BUNA N

FLOW CELL RATINGS

TEMPERATURE: 32-122F (50C)

MAX. PRESSURE: 65 PSIG (549 kPa [abs])

FLOW RATE: 2-5 GPH (7.6-18.9 LPH)

The see-through acrylic Low Flow Cell

MILLIMETER

INCH

FIGURE 2-7. Jet Spray Cleaner PN 12707-00

7

Image 11
Contents Combination pH Sensor Essential Instructions Hazardous Area InstallationModel TF396 List of Tables List of FiguresNumber Title Minimum Process Conductivity 75 µS/cm nominal 100 µS/cm Section Description and SpecificationsFeatures and Applications Ordering Information Accessories First Time Installation GuideFlow Through and Insertion Mounting Section InstallationUnpacking and Inspection Mounting4TF39604 B 40396P05 4TF39603 LOW Flow Cell PN Jet Spray Cleaner PNSection Wiring Model TF396 41055102 40508114 Sensor Preparation Instrument Configuration for USE with Model TF396Recommended two-point buffer calibration procedure Section Start UP and CalibrationRo and R1 Values for Temperature Compensation Elements Section MaintenanceAutomatic Temperature Compensator Electrode CleaningGeneral Warranty RepairNON-WARRANTY Repair Section Return of MaterialSensor or Circuit Board only Return of Materials RequestReturn of Material WarrantySpecifications subject to change without notice