OPERATION MANUAL FOR
MICROPHONE3 Series 103A
Refer to Guide
NOTE: The internal
Proper
At PCB, the units are subjected to various calibrated steps of pneumatic pressure to obtain a linearity curve over the full range of the instruments.
6.0MAINTENANCE AND REPAIR
The small size and sealed construction of the Series 103A precludes field maintenance and repair.
Should the time constant degrade or an abnormality appear in the normal bias voltage, bake the unit out in a +250° F oven for 1 to 2 hours, then
If this is not effective or if other problems should occur, contact the factory for assistance in tracing the problem or for instructions on returning the unit for repair.
4.0OUTPUT SIGNAL
The output signal from the 103A Series is a voltage proportional to input pressure. The polarity is
The signal is of high enough level and low enough output impedance to be fed directly into most oscilloscopes, tape recorders, strip chart recorders, A to D converters, etc.
Consult the factory for information regarding galvo driving since special amplifiers are needed to drive most
5.0CALIBRATION
The 103A Series Sensors must be dynamically calibrated using a known step function or pulse, or with a microphone calibrator.
Drawing Number: 21082
Revision: NR
7.0PRECAUTIONS
l.) Do not apply voltage to the unit without a
2.) Do not subject these units to temperatures higher than 250° F.
3.) Do not attempt to disassemble these units as to do so voids the warranty.
4.) Use caution when removing units that have been mounted with epoxies or other strong adhesives. Excessive squeezing or torquing may damage the units and/or cause a calibration shift.
5.) In vibratory environments, secure cables and leads to surrounding structures to avoid lead or cable damage.
6.) Do not over torque when using the Servo clamps or the clamp nut since this could damage the units.
7.) Do not overpressure (see specs for maximum pressure), to do so may change the calibration.
®ICP is a registered trademark of PCB Piezotronics