Omega Engineering CL531 manual EZ-CheckSwitch, Source Mode, Wire Transmitter Simulation Mode

Page 4

EZ-CheckSwitch

The EZ-Check™ switch has three positions -- high, dial, and low. Its position is shown at the left edge of the display with “HI” and “LO” indicators. Neither indicator indicates the middle position. Use of the EZ-Check™ switch depends on mode.

Source Modes:

Slide the EZ-Check™ switch to the HI and LO positions to recall the preset settings (Hi=20.00mA & Lo=4.00mA).

Hint: For faster calibrations, the position of the switch can be felt. This feature allows continuous monitoring of the device being calibrated without looking back at the CL531 display. This is also useful in poor lighting or under difficult operating conditions.

Read Modes:

In read mode, the CL531 calibrator records the maximum and minimum readings observed in each mode. Slide the EZ- Check™ switch to the Dial position to read the loop. Then Slide the EZ-Check™ switch to the HI and LO positions to display the max and min readings. Press EZ-Dialknobto clear the readings. The display will flash “CLEARED” to confirm.

Source Mode

Transmitter Input

(Sensor)

+ IN - REF +OUT-

 

(Process Signal)

 

(Simulated Input)

Typical

 

 

2-Wire

 

Transmitter

CL531

mA LOOP CALIBRATOR

Receiver

Power Supply 0 to 60 VDC

Source mode uses internal power to supply current from 0.00-24.00 mA into as much as 1200Ω until the end of battery life. The calibrator Graph will flash if connected improperly. The three-position EZ-Check switch provides instant preset 4mA at zero, 12mA at mid range and 20mA at full scale outputs. The output is adjusted in 0.01 or 0.10 mA increments (0.1 or 1.0) % display units with the EZ-Dial knob.

2-Wire Transmitter Simulation Mode

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Transm itter Input

 

 

 

 

 

 

 

 

 

 

 

(Sensor)

 

 

+ IN -

REF

+OUT-

 

 

 

 

 

 

 

 

 

 

 

 

 

(Process Signal)

 

 

 

 

 

 

 

 

 

 

 

(Sim ulated Input)

 

 

Typical

 

 

 

 

 

 

 

 

 

 

 

 

 

2-W ire

 

 

 

 

 

 

 

 

 

Transm itter

 

CL531

mA LOOP CALIBRATOR

Receiver

Power Supply 0 to 60 VDC

The CL531 can simulate a 2-wire transmitter in the 4-20 mA or % process loop. In source mode press the EZ-Dial™ Knob twice to get into the feature options. Then press the EZ-Dial™ Knob to select mA 2 – wire. The EZ-Check switch and EZ-Dial knob allow rapid and fine control of loop current.

2

Image 4
Contents User’s Guide Servicing North America Error Conditions Basic Keypad OperationsEZ-Dial Knob Quick Reference Bar GraphSource Mode EZ-CheckSwitchWire Transmitter Simulation Mode Power and Measure Transmitter Mode Read ModeSpecifications General SpecificationsCL531 Field Calibration Procedure Page WARRANTY/DISCLAIMER Shop online at omega.com

CL531 specifications

Omega Engineering is renowned for its precision instruments and innovative solutions in the field of measurement and control. Among its notable offerings is the Omega CL531, an advanced process controller that has gained recognition for its reliability and versatility in a wide range of industrial applications.

The Omega CL531 features a powerful microprocessor-based design, which allows for precise control and monitoring of various parameters such as temperature, pressure, and flow. This capability makes it ideal for use in laboratories, manufacturing processes, and environmental monitoring. The device features a dual-display that simultaneously shows the process variable and the setpoint, providing users with clear visibility of critical data.

One of the standout technologies integrated into the CL531 is its user-friendly interface. The controller is designed with a simple navigation system, allowing operators to configure settings and adjust parameters quickly. This ease of use is complemented by a comprehensive manual and support resources, ensuring that users can efficiently operate the device without extensive training.

The CL531 also boasts advanced control capabilities, including PID control, which optimizes the regulation of process variables. This functionality is crucial for maintaining stability in processes where precise control is essential. Users can easily adjust the PID settings to tailor the controller's performance to their specific application requirements.

In addition to its control features, the Omega CL531 supports a wide range of input types, including thermocouples, RTDs, and voltage or current signals. This versatility enables users to integrate the controller into existing systems without the need for significant modifications, making it a cost-effective solution for upgrading process control technologies.

Another significant characteristic of the CL531 is its robust construction, designed to withstand the demands of industrial environments. With its durable housing and reliable components, the device ensures long-term performance even in challenging conditions.

In summary, the Omega CL531 serves as a powerful and versatile tool for process control in various applications. Its advanced features, user-friendly interface, and robust design make it an essential instrument for industries requiring precision monitoring and control. Whether in laboratory settings or manufacturing lines, the Omega CL531 stands out as a top choice for engineers and technicians seeking reliable process management solutions.