CDE681 Part 2 - Installation

Installation Tip!

Re-using Compression Fitting, Nut, and Crimped Ferrule:

If the sensor is re-installed, the compression fitting and nut can be re-used. The crimped ferrule, however, may need to be cut away from the sensor to remove it, making it unusable. If you can remove the crimped ferrule without destroying it, and if it can still provide an effective process seal, you can probably re-use it at least the number of times listed below:

Recommended Crimped Ferrule Re-use

Ferrule Material

Number of Re-uses

316 stainless steel

1 after initial use

Kynai (PVDF)

3 after initial use

2.2Immersion Mounting

1.Reverse the compression fitting assembly on the sensor so that the longer threaded section faces towards the sensor cable end (see Figure 2-4).

2.Position the compression fitting assembly at the far cable end of the sensor body and tighten the compression nut onto the compression fitting. Use one wrench to hold the fitting and another to turn the nut 1 to 1% turns to crimp the ferrule. This provides a process seal.

f

! L---3/4OR1NCHIPIPE (CUSTOMERIDED) PROV

-:.

+- + 3/4OR1NCHIPTCOUPLING (CUSTOMERIDED) PROV

I /

I I

DIMENSIONmm(I”.)

I1

2l/NCHIX3,4INCH OR3/4INCHX1NCHI

NPT REDUCER

88.9O139T.7

(3.50O5.50)T2l/NCHIX3/4INCH

SENSORCOMPRESSIONFITTING EXTENSION

SENSOR

Figure 2-4. Immersion Mounting Details

I_

.:

Image 12
Contents CDE68 1 Series Omega .com@ EGAOM!Y-fHelpful Identifiers CDE68 1 Series Conductivity/Resistivity SensorsTable of Contents List of Figures 17/03 214 PM Pageiv Operating Procedures Part ONE Introduction General InformationCDE681 Part 1 Introduction DescriptionSpecifications Model NoConstant CDE681 -D-SPart TWO Installation Location Requirements CDE681 Part 2 lnstallatioInsertion Mounting Comtxession Avoid leaks. Recommendation UseTeflon tape or pipeDireitly in the process Flow path This completes the insertion mountingRe-using Compression Fitting, Nut, and Crimped Ferrule Immersion MountingKynai Pvdf Fasten a Unilet junction box onto the other end of the pipe M3594-0603-CDE681Series.qxd l/17/03 214 PMSENSOR/INTERCONNECT Cable Details Sensor Cable DetailsInterconnect Cable Details Shield foilConnecting Interconnect Cable Do not fold back the cellophane binder exposed in stepCDE681 Part 3 Service and Maintenance Checking Sensor Operation Measurement PointsCustomer Assistance $, +,y*$ ,,$$&ww+~~~~~~~~~~~.-‘a*‘i%*,,%*dd%q,’+b Shop online at Www,omega.com PH/CONDUCTIVITY