PART TWO - INSTALLATION

SECTION 1 - LOCATION REQUIREMENTS

Locate the sensor as close as possible to the measuring instrument. Do not exceed a distance of 91 m (300 feet) between the sensor and instrument.

SECTION 2 - Mounting

The CDE681-series compression fitting style sensor may be insertion mounted into a pipe tee or vessel fitting. By reversing the compression fitting, the sensor can be fastened to the end of an appropriate length pipe for immersion

mounting. The “longer version ” compression style sensor is intended to be mounted into a ball valve assembly, enabling sensor removal without stopping the process flow. Figure 2-l shows the sensor ’s general dimensions.

.

f

CLEAR SHIELD)W/BLACK(SEN

BLACK(SENSE)

RED(DRIVE)

WHITE(TEMP+)

BLUE(TEMP-)

CLEAR(INNER SHIELD

l/2’DIA. FOR

00.5CELL CONSTANT

3/4’DIA. FOR ALL 1

OTHER CELLl/2’OR3/4’CONSTANTS\

I,?OR3,COMPRESSION-V FIrrING

NPT THREADS

Figure 2-l. General Dimensions and Cable Wire Details

2. 1 Insertion Mounting

To ensure optimum measurement performance, follow these guidelines when insertion mounting the sensor:

??Install the sensor into the pipe run so that the process flows directly into the end of the sensor (see Figure 2-2).

??Preferably, mount the sensor in a vertical position to eliminate the possibility of trapped air bubbles from contacting its electrodes which can cause measurement error. This also prevents loose pipe line sediment from accumulating and obstructing the sensor electrodes.

3

Image 9
Contents CDE68 1 Series EGAOM!Y-f Omega .com@CDE68 1 Series Conductivity/Resistivity Sensors Helpful IdentifiersTable of Contents List of Figures 17/03 214 PM Pageiv Part ONE Introduction General Information Operating ProceduresCDE681 Part 1 Introduction DescriptionModel No SpecificationsConstant CDE681 -D-SPart TWO Installation Location Requirements CDE681 Part 2 lnstallatioInsertion Mounting Avoid leaks. Recommendation UseTeflon tape or pipe ComtxessionThis completes the insertion mounting Direitly in the process Flow pathRe-using Compression Fitting, Nut, and Crimped Ferrule Immersion MountingKynai Pvdf M3594-0603-CDE681Series.qxd l/17/03 214 PM Fasten a Unilet junction box onto the other end of the pipeSensor Cable Details SENSOR/INTERCONNECT Cable DetailsInterconnect Cable Details Shield foilDo not fold back the cellophane binder exposed in step Connecting Interconnect CableCDE681 Part 3 Service and Maintenance Measurement Points Checking Sensor OperationCustomer Assistance $, +,y*$ ,,$$&ww+~~~~~~~~~~~.-‘a*‘i%*,,%*dd%q,’+b PH/CONDUCTIVITY Shop online at Www,omega.com