Fluke PN 650454 instruction sheet

Page 10

Electromagnetic Compatibility in RF field of 3 V/m on all functions except MΩ

Relative Humidity except 32 MΩ Range:

32 MΩ Range only:

Altitude

Battery Type

Battery Life

Continuity Beeper

Shock, Vibration

Size (H x W x L)

Weight

Safety

EMC Regulations

Certifications/Listings

Total accuracy = Specified accuracy plus +2.0% of range for all functions except:

320.0mV range: total accuracy = specified accuracy + 1.0% of range

320.0Ω range: total accuracy = specified accuracy + 8% of range

0% to 90% (0°C to 35°C)

0% to 70% (35°C to 50°C)

0% to 80% (0°C to 35°C)

0% to 70% (35°C to 50°C)

Operating: 2000 meters

Storage: 12,000 meters

9 V, NEDA 1604 or 6F22 or 006P, or NEDA 1604A or 6LR61

2000 hrs typical with alkaline

1600 hrs typical with carbon zinc

4096 Hz

per MIL-T-PRF 28800F Class III, Sinusoidal, Non Operating

2.8cm x 7.5 cm x 16.6 cm (1.12 in x 2.95 in x 6.55 in)

340 g (12.0 oz)

600 V CAT II per ANSI/ISA S82.01- 1994, EN 61010-1: 1993, CSA C22.2 No 1010.1-92, UL 3111-1.

EN 61326-1: 1997

Image 10
Contents 70/73 Autorange Input JacksPushbutton Manual RangeAutomatic Touch Hold Mode Bar GraphAC and DC Voltage K L mL StandbyResistance e Diode Test G Continuity Test R Current ? aMaintenance HolsterInternal Fuse Test Specifications CleaningPage Function Overload protection for all functions and ranges 600