Instrument Configuration
The following commands are all available from the Configuration menu in the WinLog97 software installed on your PC.
The instrument must be connected to your PC in order to change its configuration.
INSTRUMENT - PC CONNECTION
The Serial port menu lets you choose one of four serial communication ports. Usually, COM1 is used to connect to a mouse, so try COM2 first.
SETTING SAMPLING INTERVALS
The analyzer can perform as a standalone data acquisition device, automatically recording gas measurements with the date and time, and storing up to 500 of these values.
Choosing Sampling Rate lets you select time intervals (acquisition rate) for this storage capability.
MEMBRANE SELECTION
You may find it necessary to use a different type of membrane for different applications.
To
CALIBRATION MODE
Use the Calibration mode command to select how the sensor is to be calibrated.
Choose In Air to allow calibration of the sensor in air, In Line to calibrate the sensor directly in the sampling line, or In a Span Gas to calibrate in a source of gas of known concentration.
CALIBRATION KEY STATUS
You can use the Calibration Key Status menu to prevent an accidental sensor
Enable or Disable the instrument front panel CAL button. Disabling the button will prevent accidental sensor
Page 6