HP 11793A manual

Page 40

Microwave Converter

Model 11793A

1.Place the RF Module upside down with the semi-rigid coax cables pointing to the left. Position the main deck over the module with the four holes in the deck aligning with the four screw holes in the module base. (The A1 assembly mounting stand-offs on MP13 (main deck) should be upside down and on the opposite side of MP 13 from the assembler.) Attach with four screws.

2.Take the main deck assembly (coax cables facing to the left) and slide into the main chassis through the front frame (the two notches indicate the bottom edge and should be down). When the coax cables are within two inches of the front frame edge, carefully connect the semi-rigid coax cables to their respective front-panel connectors, using a S/16 inch open-end wrench.

For Option 020 and 021 instruments, main deck reassembly is just the opposite of disassembly.

Be careful that you do not bend the coax cables. Possible attenuation or interference of the RF signal may result.

3.Carefully slide the main deck assembly into the instrument chassis until the front panel is within the front frame and the mounting holes are aligned. Attach the front panel with the six Torxhead screws: 3 top, 3 bottom.

4.Attach the main deck assembly to the side struts using the six screws: three screws to a side.

5. Turn the instrument over. With the three A1 assembly mounting stand-offs on the main deck (MP13) adjacent to the assembler, mount the A1 assembly to the main deck assembly. The three large holes in the A1 assembly must align with the three mounting stand-offs, and the heat sink holes must align with the deck mounting holes. Firmly seat the A1 assembly on the mounting stand-offs and lock the stand-offs by turning the screw heads a 1/4 turn clockwise.

6.Fasten the transistor heat sink to the deck with two screws.

7.Connect W12 to AlJ1.

8.Mount transformer to the mounting bracket using the four long transformer screws. Mount the bracket to the left side strut at the third and eighth holes (counting forward from the rear frame) with four screws. Make sure that the transformer secondary wires (4, 5, 6) face the corner of the A1 Switch Control Assembly.

9.Replace rear frame assembly to the side struts by connecting with four screws at the frame corners. On instruments with serial number prefix 2407A, mount the rear panel to the rear frame using eight screws. Ensure that the panel is aligned so that the electrical hardware is in the same corner as the transformer.

10.Connect the black/green primary wire (05) on VOLTAGE lug (underneath the white/brown/gray

(0)to the upper right LINE VOLTAGE lug.

the transformer to the bottom right LINE wire (9181). Connect the black primary wire

11.Solder the colored secondary wires from the transformer to their respective solder pads on the A1 Assembly: yellow to 4, green to 5: and blue to 6. Solder the black wire to the GND pad.

12.Reconnect J8 (on W 1S j to the rear panel (in the FREQ OFFSET INPUT opening) using the 9/ 16 inch lock washer and nut.

30

Image 40
Contents HP 11793A Warranty AgiIent TechnoIogies Microwave Converter Regulatory Information Safety Considerations Safety Considerationsfor this Instrument Product Markings Safety Considerations Microwave Converter Specifications General InformationPage HP 11 793A Option Guide Microwave Converter Model 11793A Model 11793A Microwave Converter Power Requirements Power CablesMating Connectors Line Voltage SelectionI0 0.J Operating Environment Bench OperationRack Mounting Storage and ShipmentOperation Operating InstructionsOperator Maintenance Panel FeaturesOperator Checks Front-Panel Features Performance Tests Description EquipmentReplaceable Parts List AdjustmentsReplaceable Parts Page Model 11793A Replaceable Parts Model 11793A Replaceable PartsReplaceable Parts Model 1 1793A Code List of Manufacturers Cabinet Parts Illustrated Parts Breakdown for Standard and Option Microwave Converter Model Service Pin Troubleshooting GeneralDevice Repair Cleanliness Cleaning the Control-Line Switch ContactsInstrument Disassembly Procedure Instrument Assembly Procedure Page Page Page Page Page Model 11793A Microwave Converter Microwave Signal DOWN-CONVERSION L Ill Ss2 Ss3 Regional Sales and Service Offices