Table 3-4 Selecting a Test Set When Multiple Units are Configured

Test Set IF Switching

A test set is selected via the built-in capability of the analyzer to generate an addressed

 

command to each test set. Each time you change the {HP-IB ADDRESS} {TEST SET}

 

function (refer to LOCAL [MENU] in the HP 8510C Operation and Service Manual), the

 

network analyzer does the following:

 

• switches the IF signal of the previously addressed test set to external

 

• switches the IF signal of the newly addressed test set to internal

 

• sets the test set’s front panel ACTIVE LED to indicate test-set status

 

• applies the active test set’s IF signals directly to J11 TEST SET INTERCONNECT

 

• passes the inactive test set’s IF signals at J10 through to J11 and on to the next

 

test set or to the network analyzer

 

 

Test Set Addressing

You can change the test set address via program control, or via the network analyzer front

 

panel keys. You must also change the mechanical switches on the test-set’s rear panel.

 

Here’s how:

 

1. Change the test set address via the network analyzer keys;

 

a. On the analyzer’s front panel, press the INSTRUMENT STATE [LOCAL] key, then

 

press {TEST SET}. The test set’s address appears on the display.

 

b. Using the key pad, enter the desired address for the test set, then press [x1]. The new

 

HP-IB address appears on the display.

 

2. Change the test set address via the HP-IB command:

 

From the controller, enter the HP-IB ADDRESS; command, followed by the

 

address value. Refer to the command reference for programming details.

 

3. Reset the mechanical switches on the test-set’s rear panel. Set the switches to match the

 

new HP-IB address displayed on the network analyzer screen.

 

 

RF Switch Driver Control

When the {HP-IB ADDRESS} {TEST SET} function is changed, the network analyzer

 

generates a code sequence. The code sequence is automatically issued across the

 

HP 8510C system bus to the device at the ADDRESS of RF SWITCH .

 

In the recommended configuration, the device is an HP 11713A attenuator/switch driver. The

 

switch/driver in turn controls one or more coaxial switches. These switches, (shown in a

 

Figure 3-3,and Figure 3-4)are used to choose which test set receives the RF output from the

 

network analyzer source.

 

The exact command issued depends upon the new value of the {HP-IB ADDRESS}

 

{TEST SET} value, also shown in shown in a Figure 3-3and Figure 3-4.

 

 

HP 8517B S-Parameter Test Set Manual 3-6

Image 44
Contents HP 8517B Parameter Test Set Restricted Rights Legend What You’ll Find in This Manual… Warranty Certification WarrantyAssistance Service and Support Safety and Regulatory Information Instrument MarkingsSafety Earth Ground Typeface Conventions ItalicsHP 8517B S-Parameter Test Set Manual Contents Installation Specifications HP 8517B Test Set Replacement Procedures HP 8517B Replaceable Parts Contents-6HP 8517B S-Parameter Test Set Manual How to Use This Manual Getting StartedTest Set Description Test Set DescriptionOption 004, HP 8517B Test Set Block Diagram HP 8517B Test Set Options HP 8517B Options and DescriptionsHP 8517B Test Set Options Options HP 8517B Test Set Option DescriptionVerifying Test Set Operation Verifying Test Set OperationTest Set Operation Verification Maintaining Measurement Accuracy Maintaining Measurement AccuracyInstruments Supported by This Manual Instruments Supported by This ManualInstrument Firmware Compatibility Source Firmware Revision Compatibility MatrixService and Support Options Available Service and Support OptionsService and Support Options Available Option DescriptionAccessories Accessories AvailableHP 8517B Test Set Accessories Available AccessoriesAccessory Model Number Contents or Description Critical Recommended Model Use Specifications Recommended Test EquipmentRecommended Test Equipment Recommended Test EquipmentOperating and Safety Precautions Safety PrecautionsMaximum Input Power Levels Operating and Safety Precautions14HP 8517B S-Parameter Test Set Manual 2Installation Initial Inspection Initial InspectionRF Source Cable, Bench-top Configuration Anti-rotation Clamps RF Source Cable, Rack ConfigurationInstalling the Test Set Into a System Rack Installing the Test Set on a BenchRecommended Static-Free Workstation Configuration Installing the Test Set Into a System RackConfiguring the Test Set in a System Configuring the Test Set in a SystemReplacing O-Rings in Anti-Rotation Clamps Connecting the System CablesConnection Instructions and Connector Descriptions Packaging the Test Set Packaging the Test Set8HP 8517B S-Parameter Test Set Manual Front-Panel Features Front-Panel DescriptionRear-Panel Features Port 1 Bias Controlling Multiple Test SetsActive devices under test HP 8517B S-Parameter Test Set Manual Initialization at Power-Up Selecting a Test Set Multiple Test-Set ConnectionsSelecting a Test Set When Multiple Units are Configured With Switch #1 With Switch #2 New AddressMeasurement Calibration Measuring High-Power Devices with Option Changing Signal Path States After System Calibration Changing the Test Port AttenuatorsStoring Trace Memories Viewing Normalized Parameters Default to Memory Making Operational Checks Instrument State Local Auxiliary Menus SystemPerformance Verification Using Anti-Rotation Clamps Attach the first clampUsing a Torque Wrench Correctly Positioning the thumb-screwPositioning the connector Visually Aligning Clamp and Nut FlatsAttaching the second clamp Using Anti-Rotation ClampsAligning the thumb-screw HP 8517B S-Parameter Test Set Manual Mechanical Specifications Supplemental CharacteristicsHP 8510/HP 8517B Mechanical Specifications HP 8510/HP 8517B Supplemental Characteristics3HP 8517B S-Parameter Test Set Manual HP 8517B S-Parameter Test Set Manual Troubleshooting the Test Set Theory of Operation RF Source PowerTheory of Operation Theory of Operation LO Signal Control Troubleshooting Sequence Troubleshooting SequenceEquipment Type HP Part Number Equipment Needed But Not SuppliedEquipment Required, But Not Supplied Troubleshooting Procedures AssemblyPower Supply Voltages to A15 Power Supply Voltages to A27Power Supply Fuses and Test Points Instrument HP-IB Switch Setting Fuse LocationIf the Self-Test Fails to Run Properly Active LED LocationSelf-Test Indication LSN TLK SRQ REMUsing the Service Adapter Service Adapter ConnectionsService Adapter Conclusions HP 85102 is working properly. The problem isResults Observations Most likely in the source or the test set. Refer toTesting Signal Paths Isolating Signal Path ProblemsUnratioed Power Test Drive Port Testing Paths 5 Test switch to restore factory default conditions10 RF Path 1 and Path 2, HP 8517B Standard Test Set 11 RF Path 3 and Path 4, HP 8517B Standard Test Set 12 RF Path 5 and Path 6, HP 8517B Standard Test Set 13 RF Path 1 through Path 4, HP 8517B Option 007 Test Set 14 RF Signal for Path 1 through Path 4, HP 8517B Option 15 RF Path 5 and Path 6, HP 8517B Option Most Probable Failures B indicates Bad indicates Good Sample DiagnosticsTest Results for Path 1 through Path 24HP 8517B S-Parameter Test Set Manual HP 8517B Test Set Replacement Procedures Performance Tests Adjustments ESD-Safe Workstation ConfigurationEquipment Needed But Not Supplied Replacement-Procedure Tools NeededTools Used for Procedure HP Part Number HP 8517B Major Assembly Location Diagram Assembly Replacement Procedures Initial Precautions Switch/Splitter ReplacementA18 Switch/Splitter Assembly A14, A10 to A13 Frequency Converter AssemblyC1 to C4 Filter Capacitors Regulator Board Replacement Filter Capacitor ReplacementA15 Regulator Board Assembly Mm Test Port Connectors Diagram of 2.4 mm Test Port ConnectorFan Assembly Replacement Power TransformerReplacement B1 Fan AssemblyThis procedure. Disconnect the ac power cord before starting A1 Front Panel and Interface Board Assembly Front Panel Interface Board ReplacementRing Replacement Anti-Rotation Clamp O-Ring Replacement Bias Tees Replacement A7 and A9 Bias Tees Test Port AttenuatorA6 and A8 Test Port Couplers A16 or A17 Port AttenuatorsA22 Input Amplifier Assembly Input AmplifierBuffer Amplifier A23 to A26 Buffer AmplifiersPost Regulator Board Replacement A27 Post Regulator Board Assembly14HP 8517B S-Parameter Test Set Manual Introduction Rebuilt-Exchange AssembliesAssembly Fuse Switch Meter Electron Tube Recovery Diode, VaractorOrdering Information To Order Parts . . . . FastManufacturer Names and Addresses Standard Abbreviations Depletion Mode DecoderDegree Shaped holeMale, Maximum, Mega, Mi, Milli, Mode HardwareHexadecimal, Hexagon, Hexagonal HelicalOpt Optical, Option, Optional Picoampere, Power Amplifier Pad, Power DissipationMiniature MiterPico Farad, Power Factor Subminiature, A-Type Threaded ConnectorPackage PlasticSubminiature, B-Type Slip-on Connector Subminiature, C-Type Threaded ConnectorTeeth, temperature, Thickness, Time, Timed, Tooth, Typical Ambient Temperature, TantalumMultipliers, Abbreviation and Description Abbreviation Prefix MultiplePower Cable and Plug Part Numbers Part number Instrument FusesFuse Name Location Low Cost Rebuilt-Exchange Procedure Hewlett-Packard pays postage on boxes mailed in the U.S.AMajor Assembly Reference Designator and Parts Information DescriptionMajor Assembly Locations in the Standard HP 8517B Test Set Cable Assembly AT1 to A10J2 Cable Assembly A8J3 to A11J2Cable Assembly A6J3 to A12J2 Cable Assembly A8J2 to A9J2Flexible Cable Assemblies in the Standard HP 8517B Test Set 18HP 8517B S-Parameter Test Set Manual Miscellaneous Parts, Front Panel HP 8517B Test Set 10 Test Port Connector Assembly Components 11 Miscellaneous Parts, Rear Panel HP 8517B Test Set Teflon Cable Sleeve Line Module Retainer ClipsWasher-LK Intl T 3/8 in 0.377-IN-ID Washer-LK Intl T 7/16 in 0.439-IN-IDDetailed Views of Hardware Stack-Ups 12 Miscellaneous Parts, Top Internal View C1 C4 0180-2671 13 Miscellaneous Parts, Motherboard2200-0107 Machine Screws Connector 5-PIN M Post Type Connector Post Type 0.100-PIN-SPCG 14-CONTConnector 5-PIN M Post Type J8 and J9 Connector-PC Edge 2-Rows XA5 XA6 to XA14 Not Assigned XA1514 Parts Unique to HP 8517B Option 001 Test Set Cable Assembly, A2J9 to J10A4 Cable Assembly, A2J6 to J10A2Cable Assembly, A2J12 to J10A3 Cable Assembly, A3J3 to J10A715 Parts Unique to HP 8417B Option 002 Test Set Cable Assembly, A12J3 to Port 1 J13Cable Assembly, A13J3 to Port 2J14 Cable Assembly, AT7 to A9J1 Cable Assembly, A8J3 to A17J2Cable Assembly, A17J1 to A11J2 16 Parts Unique to HP 8517B Option 004 Test SetPost Regulator Board Assembly Beneath A22 Input Amplifier 17 Parts Unique to HP 8517B Option 007 Test Set08516-60021 DB Attenuators Input AmplifierHP Part Number Front Panel J7 to A22 Input AmplifierA22 Input Amplifier to A18J5 A6 Coupler to A2318 Parts Unique to Option 002 plus Option 19 Parts Unique to Option 004 plus Option 20 Instrument Chassis Replaceable Parts 36HP 8517B S-Parameter Test Set Manual Index HP 8517B S-Parameter Test Set Manual Index-1Index-2HP 8517B S-Parameter Test Set Manual