independently)
In the case of DO1
Pulse rate: 3.6 to 3,600,000 pulses/h
Pulse width:Settable within the range of 0.3 to 500ms
In the case of DO2
Pulse rate: 3.6 to 360,000 pulses/h
Pulse width: Settable within the range of 4 to 500ms
•Rage switching output
One Digital Output (DO) is used
・Single direction,
Two Digital Outputs (DO) are used
・Single direction,
•High/low limit alarm output
An alarm is output when flow rate goes above or below the
Setting range :
High limit 2 points, low limit 2 points can be set At the time of alarm output, Normally Open or Normally Closed contact can be selected
• Preset point output
Contact ON when totalizer count exceeds the set value
Setting range: 1 to 99999999 count
• Converter error alarm
An alarm is output when an operation error is detected by
At the time of alarm output, Normally Open or Normally Closed contact can be selected
Communications signal:
A digital communications (HART protocol) signal is superimposed on 4 to 20 mA dc analog output signal.
Load resistance: 240 Ω to 1 kΩ
Load capacitance: 0.25 µF or less
Load inductance: 4mH or less
(For maximum cable length, about 2km is a guideline length when
Note: HART (Highway Addressable Remote
Output display:
(With back light).
2 units can be selected from the flowing units: flow velocity, instantaneous flow rate, total flow (forward/reverse/difference flow), total count, %, custom unit.
Parameter
•IR Switches: Various parameters can be set without opening the converter housing using 4 infrared switches (password can be set)
•Zero adjustment: Zero point adjustment can be started by pressing the switch in the converter.
•Digital communication: The AF 900
•Damping: 0.5 to 60 s
(Selectable in one second increments) “Field
Zero span calibration tool allows unit to be
Conditions when power fails:
Parameter setting values are stored in
Current output: 0 mA
Digital output : OFF (contact open)
LCD display : No display
Surge protection: Surge protectors are installed in the power supply, excitation circuit, current signal output and digital I/O circuit.
Terminal block
Housing: Aluminum alloy
Coating: Acrylic
Cable connection ports :
A cable gland is provided for each port.
OD of cable φ11 to 13 mm
Material Nylon 66
G(PF) 1/2 male screws.
Structure: IP67 and NEMA 4X Watertight
Vibration resistance:
No resonance to the following levels of vibration:
•10 to 55Hz with amplitude of 0.07mm,
•Vibration of 30Hz with 29.4 m/s2 in 4h in each direction will not cause any defect to unit.
Note: Avoid using the flowmeter in an environment with constant vibration.
Dimension and Weights:
See Figure 5
3