PX791 & PX792 SERIES
OMEGA® SANITARY PRESSURE TRANSDUCER
INSTRUCTION SHEET
W A R N I N G !
This instrument is susceptible to damage when exposed to static electrical charges. To avoid damage to the transducer observe the following:
•Ground the body of the transducer BEFORE making any electrical connections
•When disconnecting, remove the ground LAST.
NOTE: The braided shield and drain wire in the cable (if supplied) is not connected to the transducer body, and is not a suitable ground.
CAUTION: Pressure spikes in excess of the rated overpressure capabil- ity of the transducer may cause irreversible electrical and/or mechani- cal damage to the pressure measuring and containing element(s).
Mounting
Although the unit can withstand normal vibra- tion without damage or significant output effects, it is always good practice to mount the transducer where there is minimum vibration. Be sure to use a gasket that does not interfere with the sanitary diaphragm. If the gasket I.D. is smaller than 1.50 inches, an offset due to clamping force will occur.
Power Supply
Excitation (Ratiometric Output Only)
For proper operation a voltage within the range of 5 to 10 Vdc must be applied between the transducer’s supply terminals.
Noise
For minimum noise susceptibility, avoid running the transducer’s cable in a conduit that contains high current AC power cables. Where possible avoid running the cable near inductive equipment.
The supply voltage for the
Shield Wiring
Connect the braided shield to the guard terminal on the reading instrument (meter, etc.) if avail- able or to ground or to the power supply nega- tive terminal.
Adjustment Potentiometers
The zero and span pots are accessible through the top of the case. Loosen the four screws and separate the top carefully. The zero pot is marked with a white dot.