Fluke Ti20 user manual Setting Alarm Limits

Page 45

Advanced Imager Operation 3

Setting Alarm Limits

4.In the Adjust RTC function, Press F(2) to increase the RTC value or H(1) to decrease the RTC value.

dag122f.bmp

5. Press G(DONE) to return to the Home display.

Setting Alarm Limits

When you adjust alarm limits, the lower alarm limit appears in blue text in the information zone and upper alarm limit appears in red text in the information zone. The low alarm indicator () and high alarm indicator () also move on the temperature scale.

When you scan a target that registers a temperature exceeding one of the alarm limits:

A blinking alarm indicator appears on the Imager display

If the center pixel target temperature exceeds the alarm limit, the center image will blink and alarm value turns bold blue or bold red depending on which limit has been exceeded.

The default low alarm limit is -10 °C (14 °F) and the default high alarm limit is

350°C (662 °F).

1.From the Home display, press G(MENU) four times to navigate to the Adjust Alarm and Sleep Mode display.

3-7

Image 45
Contents Ti20 Limited Warranty and Limitation of Liability Table of Contents Ti20 List of Tables Ti20 List of Figures Ti20 Getting Started IntroductionContacting Fluke Symbols Safety InformationLaser Warning Labels Ti20Unpacking the Imager Standard AccessoriesUnpacking the Imager Ti20 Features and ControlsOperating the Controls Focusing the ImagerUnderstanding the Trigger Ti20Using the AC Power Adapter Using the AC Power AdapterCharging and Replacing the Batteries Using the Battery ChargerReplacing the Battery Pack Installing or Replacing the BatteriesAttaching the Wrist Strap Attaching the Wrist StrapInputs and Connections Connecting the USB CableCleaning Cleaning the CaseMounting the Imager on a Tripod Cleaning the Lens Ti20 Turning the Imager On and Off Basic OperationImager Home Display Zones Understanding the Home DisplayContents of the Home display is described in Table Aiming and Activating the Laser Comparing Frozen Images to Stored Images Capturing ImagesAdjusting the Backlight Setting the Level Setting the Temperature ScalePress Flevel to access the set Level function Adjusting the Span Using Distance to Spot Size Ratio DSManually Activating the Calibration Flag FOV Environmental Conditions Ambient Temperature Derating and Thermal ShockEmissivity Tape methodContact thermometer method Reflected Temperature Compensation Reflected Temperature CompensationTi20 Advanced Imager Operation Data Management and StorageViewing Stored Images Deleting Images Selecting a Palette Adjusting Emissivity Press FNto access the Adjust Emissivity display Adjusting Reflected Temperature Compensation Values Setting Alarm Limits Dag124f.bmp Adjusting Sleep Mode Ti20 Appendices Page Glossary Blackbody Display ResolutionCalibration Calibration SourceFocus Point or Distance Field of View FOVFull Scale Full Scale AccuracyLevel Optical ResolutionMicron or µm Minimum spot sizeResolution Reflected Temperature CompensationRelative Humidity RepeatabilityTemperature Storage Temperature RangeTemperature Coefficient or Ambient Derating Temperature ResolutionWhy use non-contact infrared thermometers? Basics of Infrared MeasurementHow does IR work? What is emissivity? How to assure accurate temperature measurement?What is the distance to spot ratio? How to take temperature measurement?What are some uses for non-contact thermometers? Ti20 Typical Emissivity Values Table C-1. Emissivity Values for Metals Material To 14 µm Table C-1. Emissivity Values for Metals Material To14 µm Tin Unoxidized 0.3 Titanium Polished Typical Emissivity Values Ti20 Thermal SpecificationsOther