Sanyo MCO-19AIC instruction manual Lighting the UV lamp for 24 hours, 120

Page 30

ξThe UV timer can be set from 0 to 30 min. The default setting is for 5 min. When the UV timer is set to zero, the UV lamp will not light.

ξIf the outer door is opened while the UV lamp is lit, the lamp will turn OFF. Then, when the door is closed, the lamp will light for the preset period of time.

ξIf the UV lamp ON time is set for longer than 10 minutes or if only the outer door is repeated opened and closed, condensation may occur in the chamber and may affect temperature distribution. It will also shorten the service life of the UV lamp.

ξWhen replacing the UV lamp, consult with a Sanyo representative or agent.

Note: To compensate for the drop in UV ray output along with increased accumulated ON time of UV lamp, the Incubator automatically extends the UV lamp ON time according to the accumulated ON time. The automatically extended time is displayed as a percentage under UV Timer Ext. in the UV Setting Screen.

Lighting the UV lamp for 24 hours

If the chamber has been contaminated by dirt or by spilling the medium, use the following procedure to decontaminate the chamber by lighting the UV lamp for 24 hours.

1.Remove all attachments from the chamber, including the trays, tray supports, side supports, fan cover, duct, fan, humidifying pan, and humidifying pan cover. Clean all the attachments in an autoclave or with disinfectant alcohol.

2.Clean and wipe off inside the chamber with disinfectant alcohol.

3.Set the CO2 density to 0% and set the UV 24h Mode to 1 on the UV Setting Display.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

U V

S e t

t i

n g

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M E N U

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3

U V

T i m e r

 

5

m i n ( 0

- 3 0

m i n )

 

 

 

 

O

K

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

U V

L i f

e

 

0

%

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

C a n c e l

 

 

4

U V

T i m e r

E x t .

+ 0

%

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

U V

2 4 h

M o d e

 

1

 

( 0 . O F F

1 . O N )

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4. Press the MENU Key to display the menu, select OK, and press the ENTER Key. The parameter will be saved.

This procedure must be preformed only with the outer door closed and the UV lamp turned OFF.

The UV lamp will light continuously for 24 hours after this parameter is set. The setting is canceled if the outer door is opened. Perform the above procedure to set the 24-hours lighting mode again if you open the door.

Install the attachments after completion of the 24-hour lighting mode.

Note: Set the upper limit alarm temperature to at least 10oC higher than the chamber set temperature when using the 24-hour lighting mode of UV lamp. The 24-hour lighting mode may cause an automatic chamber temperature alarm because the temperature of the chamber will increase.

29

 

120

 

Image 30
Contents MCO-19AIC㩿UV㪀MCO-19AIC Contents Introduction Precautions for Safe Operation Unqualified personnel may cause electric shock or fire Abnormal operation may cause electric shock or fire Page Labels on the Incubator Emvironmental ConditionsIncubator Components 998Remote alarm terminals Refer to 100Control panel and keypad 101Remote alarm terminals 102Installation Site 103Installation 104When the MCO-21GC is mounted Connecting a CO2 gas cylinderWhen the MCO-21GC is not mounted 105Preventing Contamination 106Precautions for Cultures 107Correct Operation Using the Unlock KeyUnlocking when power is interrupted 108109 Basic Operations on Control PANEL㩷 110Setting the chamber temperature and CO2 density㩷 Basic Parameters111 Setting the key lock㩷 Removing the key lock㩷112 Setting the key lock password 113114 Setting the upper limit alarm temperature 115Alarm PARAMETERS㩷 116Using the UV lamp㩷 UV Lamp Parameters117 Precautions when using the UV lamp Be careful when handling the UV lamp118 Setting the UV lamp on period 119Lighting the UV lamp for 24 hours 120Setting the date, time, and log interval 121122 Initial settings LCD/DAQ parameters 123Displaying the LOG㩷 124Transferring data 125Water Level Sensor 126Routine Maintenance Cleaning the chamber and Inner attachments127 128 129 Filling the humidifying pan 130H2O2 Decontamination 131132 H2O2 decontamination 133OKP 134135 Precautions when handling H2O2 decontamination reagent 136ALARMS, SAFETY, and SELF-DIAGNOSIS Alarms and Safety Functions for Culture Operations137 Alarm and Safety Functions for H2O2 Decontamination 138Temperature/CO2 calibration  Calibration139 140  Troubleshooting Symptom Items to check and countermeasures141 142 Disposing of the CO2 Incubator 143144 EnglishGerman 145 FrenchSpanish 146 Portuguese147 DutchSwedish Automatic CO2 Cylinder Changeover 148149 Automatic CO2 Density Calibration 150151 Menu, select OK, and press the Enter Key     152Stacking Incubators 153Front Specifications 155Performance 156Safety check sheet 157

MCO-19AIC, UV specifications

The Sanyo UV, MCO-19AIC is a state-of-the-art environmental chamber designed specifically for the storage and cultivation of microorganisms, cell cultures, and other sensitive biological samples. This advanced incubator features a unique combination of technologies that ensures precise temperature and humidity control, creating optimal conditions for various laboratory applications.

One of the standout features of the Sanyo MCO-19AIC is its advanced cooling system. The unit employs a dual air jacket design that offers superior temperature stability and uniformity. With a temperature range of 0°C to 65°C, it can accommodate a wide variety of experiments and samples. This temperature control is complemented by an integrated microprocessor that continuously monitors and adjusts the conditions inside the chamber, ensuring that they remain stable even when ambient conditions fluctuate.

Another important characteristic of the Sanyo MCO-19AIC is its humidity control capability. Utilizing an advanced humidification system, this incubator can maintain relative humidity levels from 40% to 95%. This feature is particularly important in applications requiring precise moisture levels, such as cell culture or the preservation of sensitive biological specimens.

The Sanyo MCO-19AIC is also equipped with a UV sterilization system that enhances safety and cleanliness within the chamber. This system utilizes ultraviolet light to eliminate potential contaminants, ensuring that the environment remains sterile for your sensitive samples. The use of UV light minimizes the risk of cross-contamination and provides peace of mind when working with critical biological materials.

In terms of user interface, the Sanyo MCO-19AIC features a digital display with intuitive controls that allow for easy set-up and monitoring of parameters. Users can easily adjust temperature, humidity, and other settings, while real-time data is continuously displayed on the screen. Additionally, alarms and monitoring systems alert users to any potential deviations from the set conditions, maintaining the integrity of the samples housed within.

Overall, the Sanyo UV, MCO-19AIC is an indispensable tool for any modern laboratory looking to cultivate or store biological samples with precision and reliability. Its combination of advanced cooling, humidity control, and sterilization features ensures that researchers can conduct experiments with confidence, making it a preferred choice in various fields, including microbiology, biotechnology, and pharmaceuticals. With its robust design and innovative technologies, the Sanyo MCO-19AIC stands out as a leading incubator in the scientific community.