Compur 501 IR manual Construction et fonctions Construction and Function

Page 3

2.

Construction et fonctions

2.

Construction and Function

1Support

2Presse etoupe

3Borne au sol

4Compartiment électronique avec plaque signalétique: type, n° de série, date de fabrication

5Compartiment du capteur

6Anneau pour sangle

7Protection pour éclaboussure

8Adaptateur de calibartion

9LED de service

10Affichage

11Vis bouchon

11

10

9

R

 

 

 

 

1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

1

Mounting bracket

 

 

 

 

 

 

 

 

 

 

 

2

Cable gland

 

 

 

3

Grounding terminal

 

 

 

4

Electronic compartment

3

 

 

 

 

 

 

 

with type plate,

 

 

 

 

 

 

4serial number / date code

5 Sensor compartment

5

6 Ring for holding strap

67 Splash guard

8 Calibration adapter

 

9

Service LED

7

10

Display

 

 

11

Plug screw

8

Le Statox 501 IR détecte les gaz combustibles (Méthane, Butane) et les vapeurs (Méthanol, Toluène) dans une gamme de 0 à 100 % LIE. La tête de détection est un transmetteur 4-20 mA. Il peut être utilisé avec le module de contrôle dédié Statox 501 ou avec tout autre système de contrôle de procédé.

Support

Le support et boîtier terminal sont fabriqués en polyamide renforcé de fibres conductrices. Il est de sécurité accrue pour l’utilisation dans des environnements explosifs (EExe), et est livré avec un presse étoupe (M20) et une vis bouchon de l’autre côté. Un second presse étoupe peut remplacer cette vis en option.

Boîtier du tranmetteur

Le boîtier en acier inoxydable du transmetteur se compose d’une partie supérieure et inférieure anti- déflagrantes. Il contient l’électronique, le capteur et l’affichage LED à trois chiffres. L’orifice de la sonde est protégé contre les dommages de l’eau par un filtre anti- éclaboussure.

Capteur

Le capteur est un capteur NDIR qui consiste en un source de lumière et deux photo détecteurs. La longueur d’onde de mesure est calibré sur la longueur d’onde d’absorption de la liaison C-H des hydrocarbures (3,4 µm). Plus il y a de molécules de gaz dans le faisceau de lumière (plus la concentration du gaz est grande) plus la lumière infra rouge sera absorbée. Le détecteur de référence travaille

àune longueur d’onde différente à laquelle aucune absorption n’aura lieu. Cela permet ainsi de compenser les effets de la poussière, de l’humidité ou des variations de l’intensité de la source lumineuse.

Adaptateur de calibration

L’adaptateur optionnel de calibration se branche sur la partie inférieure du détecteur. Il est équipé d’un raccord pour tube 4-6mm et de deux boutons magnétiques. Ces boutons poussoirs permettent d’accéder au menu, pour la calibration et le réglage des différents paramètres.

L’utilisation avec la sortie de gaz optionnelle permet de travailler avec un débit. Pour prévenir toute fuite, le joint ne doit pas être endommagé et toujours être fixé solidement dans sa nervure. L’adaptateur peut être fixé de manière permanente au détecteur avec deux vis.

The Statox 501 IR detects combustible gases (e.g. Methane, Butane) as well as vapors (e.g. Toluene, Methanol) in the range of 0 - 100% LEL. The sensor head is a 4 - 20 mA transmitter. It can be used in combination with the dedicated Statox 501 controller or any process control system.

Mounting bracket

The mounting bracket / terminal box is made of conductive fiber re-enforced polyamide. It features increased safety for use in explosive environments (EEx e), and comes with one cable gland (M 20) and one screw plug on the other side. An optional second cable gland can replace this.

Transmitter Housing

The stainless steel transmitter housing consists of a top and bottom part, which are rated as explosion proof. It contains the electronics, the sensor and the three-digit LED display. The sensor orifice is protected from water damage by a disposable splash guard.

Sensor

The sensor is a NDIR sensor consisting of a light source and two photo detectors. The measuring wavelength is calibrated to the absorption wavelength of the C-H bond of hydrocarbons (3,4 µm). The more gas molecules are in the way of the light beam (i.e. the higher the gas concentration is) the more infrared light will be absorbed. The reference detector works at a wavelength in which no absorption takes place to compensate for dust, humidity and variations of light source intensity.

Calibration adapter

The optional calibration adapter can be connected to the bottom side of the sensor head. It has a gas intake for a 4 x 6 mm tube and two magnetic pins. The pins operate Hall sensors allowing you to access the password-protected menu, for calibration and setting parameters.

Used with an optional gas outlet, it can serve as a gas adapter for flow applications. To avoid leaks, the integrated O-ring must be undamaged and always be fitted securely in its groove. The calibration adapter can permanently be fixed to the transmitter with 2 screws.

Image 3
Contents Statox 501 IR Statox 501 IR Sensor Head Instructions de sécurité Safety instructionsManuel ManualConstruction et fonctions Construction and Function Montage et connections Start-up and Password Menu Zero and Span Adjustment Select codeDémarrage et Menu Sensor Replacement Remplacement du capteurPAr Don Diagramme du menu / Menu diagramm100 IotISt 000Er4 Er1Er2 Er3EMV Caratéristiques techniques Technical DataUSA, Canada, Mexico Compur Monitors Inc Belgium, Netherlands, Luxemburg Compur Monitors B.VFrance Compur Monitors Sarl

501 IR specifications

The Compur 501 IR is an advanced infrared gas analyzer focused on the detection and measurement of trace gases in various applications. Renowned for its precision and reliability, the Compur 501 IR provides a comprehensive solution to industries requiring data on gases such as carbon dioxide, methane, and volatile organic compounds. This state-of-the-art instrument is vital for environmental monitoring, industrial processes, and safety compliance.

One of the standout features of the Compur 501 IR is its dual-beam infrared absorption technology. This technology enhances measurement accuracy by minimizing errors that can arise from variations in temperature and pressure. The dual-beam setup allows for continuous measurement of gas concentrations, providing real-time data crucial for process control and environmental analysis. The analyzer's sensitivity enables it to detect even the slightest changes in gas concentrations, ensuring operators are alert to any potential hazardous conditions.

The Compur 501 IR is equipped with a sophisticated microprocessor-based system that facilitates user-friendly operations. The intuitive interface allows users to configure settings, view measurements, and access calibration features with ease. Additionally, the analyzer supports various communication protocols, enabling seamless integration with existing data management systems, which is critical for efficient operation and record-keeping.

In terms of durability, the Compur 501 IR boasts robust construction designed to withstand harsh operating conditions. It features a rugged enclosure that protects sensitive components from dust, moisture, and physical impact, ensuring consistent performance in challenging environments. Moreover, the device is engineered for low maintenance, with long-lasting components that reduce the need for frequent servicing.

Another key characteristic of the Compur 501 IR is its adaptability to different calibration standards. This flexibility makes it suitable for a wide range of applications, from industrial emissions monitoring to indoor air quality assessment. The capability to customize calibration settings allows the analyzer to meet specific regulatory requirements, ensuring compliance with environmental standards.

Overall, the Compur 501 IR stands out for its innovative technology, exceptional accuracy, and user-centric design. Its ability to deliver precise gas measurements in real time is invaluable for industries focused on safety, regulation, and environmental stewardship. This advanced gas analyzer exemplifies the forefront of gas detection technology, making it a reliable choice for professionals across multiple sectors.