Selecting Fan Minimum Run Time – “Frr”
To select the Fan Minimum Runtime setting, in normal mode, press the Next button twice to get to “Con” or the Configuration menu. Then press the Enter button to enter the Con menu. The ninth selection is the “Frr” or Fan Minimum Run Time. Press Next eight times to get to “Frr” then press Enter. If the runtime is “OFF” (disabled) press Next to change it to 3, 5, 10 or 15 minutes (flashing) then press Enter to confirm the change (solid) and press Enter again to return to “run” in the Con menu. Press Next until “End” is displayed then press Enter to return to normal operation.
Selecting 4-20mA Output Option – “420”
To select the
Onboard Diagnostics
The 3M™ Macurco™
The
•0mA is most likely a connection problem
•
•24mA indicates a Trouble condition
Error Codes
•t01 Sensor is missing
•t02 Temperature compensation failed
•t04 Bad EEPROM checksum
•t08 Sensor is shorted
•t10 Bad EEPROM
•t20 Bad factory calibration
•t40 Factory calibration was not done
•t80 SDADC reading failed
If the error mode repeats frequently, check for continuous power and proper voltage. If power is not the problem and a unit has repeating error conditions, it may need to be returned to 3M Macurco for service, as noted in these User Instructions.
Sensor Poisons
The gas sensor in the detector is designed with extreme sensitivity to the environment. As a result, the sensing function may be deteriorated if it is exposed to a direct spray from aerosols such as paints, silicone vapors, etc., or to a high density of corrosive gases (such as hydrogen sulfide, sulfur dioxide) for an extended period of time.
MAINTENANCE
The 3M™ Macurco™
WWARNING
Do not disassemble unit or attempt to repair or modify any component of this instrument. This instrument contains no user serviceable parts, and substitution of components may impair intrinsic safety, which may adversely affect product performance and result in sickness or death.
CAUTION
•Avoid the use of harsh cleaning materials, abrasives and other organic solvents. Such materials may permanently scratch the surfaces and damage the display window, labels, or instrument housing.
Cleaning
Cleaning of the external surfaces is best carried out using a damp cloth with a mild detergent or soap. Vacuum under the cover with soft brush to remove dust or contamination.
Testing
WWARNING
Using a certified gas with a concentration other than the one listed for this instrument and sensor when conducting a calibration or calibration verification test (bump test) will produce inaccurate readings. This means that higher levels of the gas being monitored may be present and could result in overexposure and cause sickness or death. For proper use, see supervisor or User Instructions, or call 3M in U.S.A.,
General
There are multiple ways to test the 3M™ Macurco™
14 | 15 |