Emerson Series 3000 manual Input signals, Output signals, Non-intrinsicallysafe input signals

Page 6
Input signals

Input signals

Non-intrinsically safe input signals

 

One 2-wire

Frequency range

0–20,000 Hz

frequency/pulse input

Minimum pulse width

25 μ-sec

 

Power

Sourcing or sinking

 

Voltage

0–0.8 VDC low state

 

 

3–30 VDC high state

 

Current

Nominal 5 mA pull-up

Two discrete

Pulse width

0.15 sec minimum

momentary inputs

Voltage

0–0.8 VDC low state

 

 

3–30 VDC high state

 

 

Dry contacts

Intrinsically safe input signals

One 4-wire Coriolis sensor signal input with ground

Output signals

Non-intrinsically safe output signals

Two active

4–20 mA outputs

Three discrete outputs

One 2-wire frequency/pulse output

Isolated to ±50 VDC from all other outputs and earth ground

Maximum load limit of 1000 Ohms

Output is linear with process from 3.8 to 20.5 mA per NAMUR NE43 (June 1994)

Configurable for the application

 

Polarity

Active high or active low, software selectable

Power

Internal pull-up to 24 V

Voltage

24 VDC nominal

Current

Sourcing at 5.6 mA when Vout = 3 VDC

 

Sinking up to 500 mA at 30 VDC maximum

Scalable to 10,000 Hz

Output is linear with flow rate to 12,500 Hz

Pulse width

50% duty cycle above the crossover frequency(1)

 

Configurable between 0.543 ms and 277 ms

Polarity

Active high or active low, software selectable

Power

Active or passive, software selectable

Voltage

24

VDC nominal, active

 

30

VDC applied maximum, passive

Current

Sourcing at 10 mA at 3 VDC, active

 

Sinking at 500 mA, active or passive

(1)The crossover frequency depends on the configured value of the pulse width. At the minimum pulse width value of 0.543 ms, the crossover frequency is 922 Hz. At the maximum pulse width value of 277 ms, the crossover frequency is 1.8 Hz.

6Micro Motion® Series 3000 Transmitters and Discrete Controllers

Image 6
Contents Product Data Sheet Large operator interface for easy setup & controlWide variety of applications 2400S Micro Motion transmitters and controllers1700 2500Contents Series 3000 transmitters and discrete controllersRack-mount HousingsApplications Panel-mountPower connection Interface/displayElectrical connections DisplayNon-intrinsicallysafe input signals Input signalsOutput signals Intrinsically safe input signalsOutput signals continued Fault indicationDigital communications AC power supply Environmental specificationsPower supply DC power supplyHazardous area classifications UL and CSAATEX Mass flow accuracy12 Volume flow accuracy1 Model 3100 accessoryModel 3500/3700 liquid flow performance Relays moduleModel 3500/3700 gas flow performance WeightModel 3500/3700 density performance liquid only Sensor modelDimensions in DimensionsRack-mountModel 3300/3500 inchesDimensions continued Dimensions ininches inches Dimensions continuedDimensions in DIN-rail-mount screw terminals detailDimensions in inchesmm Field-mountModel 3350/3700Dimensions continued Conduit openings viewOrdering information - Model 2Available with Chinese M language option onlyCode Ordering information - Model 3300 continuedCode CodeOrdering information - Model Model2Available only with language option M Chinese Code Ordering information - Model 3350 continuedCode CodeOrdering information - Model Sensor interface code 6 remote core processor2Available with Chinese M language option only Code Ordering information - Model 3500 continuedCode CodeSensor interface code 6 remote core processor Ordering information - ModelModel 2Available only with language option M ChineseCode Ordering information - Model 3700 continuedCode CodeModel 1 Not available with approval code BPage Page Page Technology leadership Product breadthUnparalleled value