Agilent Technologies Detailed Specifications of the 8596E Water Dispenser

Page 5

Option Specifications

Option 010 and 011 Tracking Generator Frequency Range

8591E

100 kHz to 1.8 GHz

(Opt. 011, 75 )

1 MHz to 1.8 GHz

8593E, 94E, 95E, 96E

9 kHz to 2.9 GHz

Output Level

Range

 

8591E

0 to –70 dBm

8591E (Opt. 011)

+42.8 to –27.2 dBmV

8593E, 94E, 95E, 96E

–1 to –66 dBm

Resolution

0.1 dB

Absolute Accuracy

 

(@ 300 MHz, –20 dBm, +28.8 dBmV)

8591E

±1.0 dB

8593E, 94E, 95E, 96E

±0.75 dB

Vernier

Range

 

8591E

10 dB

8593E, 94E, 95E, 96E

9 dB

Accuracy

 

8591E

±0.75 dB

8593E, 94E, 95E, 96E

±0.5 dB

Output Attenuator

Range

 

8591E

0 to 60 dB, 10 dB steps

8593E, 94E, 95E, 96E

0 to 56 dB, 8 dB steps

Output Flatness

8591E

±1.75 dB

8593E, 94E, 95E, 96E

 

(>10 MHz)

±2.0 dB

Effective Source Match (Characteristic)

8591E

1.6:1

(10 dB attenuation)

8593E, 94E, 95E, 96E

1.5:1

(8 dB attenuation)

Spurious Output

Harmonic Spurs

 

8591E

 

(0 dBm, +42.8 dBmV output)

<–25 dBc

8593E, 94E, 95E, 96E

 

(–1 dBm Output)

 

Nonharmonic Spurs

 

8591E

<–30 dBc

8593E, 94E, 95E, 96E

≤–27 dBc

300 kHz to 2.0 GHz

2.0 GHz to 2.9 GHz

≤–23 dBc

Dynamic Range (Characteristic)

 

 

 

Dynamic Range5

TG Feedthrough

8591E

 

106 dB

≤–106 dBm

8591E

(Opt. 011)

100 dB

≤–57.24 dBmV

8593E

(> 400 kHz)

111 dB

≤–112 dBm

8594E

(> 400 kHz)

106 dB

≤–107 dBm

 

(> 5 MHz)

111 dB

≤–112 dBm

8595E

(>400 kHz)

109 dB

≤–110 dBm

8596E

(> 400 kHz)

109 dB

≤–110 dBm

Power Sweep

 

 

Range

 

 

 

8591E

(–15 dBm to 0 dBm) –(source

 

 

attenuator setting)

8591E (Opt 011)

(+27.8 to 42.8 dBmV)–(source

 

 

attenuator setting)

8593E, 94E, 95E, 96E

(–10 dBm to –1 dBm)–(source

 

 

attenuator setting)

Resolution

0.1 dB

 

Option 103 Quasi-Peak Detector

Amplitude response conforms with Publication 16 of Comité International Spécial des Perturbations Radioélectriques (CISPR) Section 1, Clause 2.

Option 105 Time Gated Spectrum Analysis

Gate Delay

Range

1

µs to 65.535 ms

Resolution

1

µs

Accuracy

±(1 µs + 0.01% x Gate Delay Readout)6

(From Gate Trigger Input to positive edge of Gate Output)

Gate Length

Range

1

µs to 65.535 ms

Resolution

1

µs

Accuracy

±(0.2 µs + (0.01% x Gate Length

 

Readout))

(From positive edge to negative edge of Gate Output)

Additional Gate Amplitude Error7

Log Scale

<2 µs

±0.8 dB

 

≥2 µs

±0.5 dB

General Specifications

Temperature Range

Operating

0°C to +55°C

Storage

–40°C to +75°C

EMI Compatibility

Conducted and radiated

 

interference CISPR Pub. 11

 

and Messempfaenger

 

Postverfuegung 526/527/79.

Audible Noise

<37.5 dBa pressure and <5.0 Bels

 

power (ISODP7779)

5.Maximum output level minus TG feedthrough.

6.Up to 1 Vs jitter due to 1 µs resolution of gate delay clock.

7.With GATE ON enabled and triggered, CW Signal, Peak Detector Mode.

5

Image 5
Contents Frequency Specifications SpecificationsFrequency Range Frequency ReferenceFrequency Readout Accuracy Resolution BandwidthMarker Count Accuracy Frequency SpanAmplitude Range Amplitude SpecificationsGain Compression Displayed Average Noise LevelResolution Bandwidth Switching Uncertainty Marker Readout Resolution 0.05 dB for log scaleSpurious Responses Reference LevelGeneral Specifications Option SpecificationsData Storage nominal Lnputs/OutputsUser Program Memory Related Literature
Related manuals
Manual 8 pages 37 Kb

8596E, 8595E, 8594E, 8593E, 8591E specifications

Agilent Technologies, a leader in test and measurement solutions, introduced the E-series spectrum analyzers—specifically the 8593E, 8594E, 8591E, 8595E, and 8596E models—designed to meet a wide range of frequency analysis needs. These models are part of Agilent's legacy of portable bench instruments that provide high performance in a compact form factor.

The Agilent 8593E is a versatile spectrum analyzer that covers frequencies from 9 kHz to 3 GHz. It features a high-resolution display and features like amplitude and frequency accuracy that make it ideal for troubleshooting and evaluating RF components. The 8594E extends the frequency range to 6 GHz, making it suitable for work in industries involving advanced telecommunications and broadcasting applications.

The 8591E is a more economical option in the lineup, operating in the same frequency range as the 8593E. It is designed for basic spectrum analysis tasks, emphasizing ease of use and practicality. In contrast, the 8595E caters to more demanding applications, offering frequency coverage up to 13.2 GHz, making it an ideal choice for emerging technologies such as 5G.

On the higher end, the 8596E model presents an extended bandwidth, covering frequencies from 9 kHz up to 22 GHz. This model includes advanced features such as built-in preamplifiers, which enhance its sensitivity, and a wide range of analysis capabilities. This allows engineers to carry out thorough investigations of both linear and non-linear devices in a single setup.

Key technologies across these models include Agilent's patented digital signal processing, which increases the accuracy and speed of frequency sweeps while minimizing noise. The instruments also boast features such as an integrated tracking generator, Applicator Simulator, and a variety of measurement applications that cater to both basic and complex analysis tasks.

In essence, the Agilent E-series spectrum analyzers are defined by their precision, versatility, and cutting-edge technology, thereby establishing themselves as essential tools for engineers and technicians across telecommunications, electronics testing, and regulatory compliance testing. Whether for fieldwork or lab environments, these analyzers are built to deliver reliable performance.