2.1 General Description

Refer to 1060-005 drawing located in Appendix A .
The Fluke Biomedical local , remote , and flasher alarms are used in conjunction with
the Victoreen Model 1060AM-NM-LR. The operating range of this system is 0.01 mR/hr
to 1R/hr. The red halogen lamps and the flashing yellow sign provides a remote (75ft)
and a local visual alarm indication. The green incandescent lights operate at half power
and indicate normal operating conditions while in a radiation field of less than 2mR/hr
+/- 10%.
An efficient dual led (red / green) operates in a similar fashion and shall alert personnel
of a bulb burn out in either the local or the remote alarms.
The Local alarm also includes two female power connectors (J1 and J2) each with a set
of normally closed contacts. J3 is a male connector with a dust cover that interfaces
with the 1060am-nm-LR . A ½ inch conduit fitting is provided for the hot (L1), neutral
(L2), and earth ground connections.
The Remote and Flasher alarm both have a single male plug connector (P1) to
interconnect with the local alarms normally closed contacts. Conduit fittings provide
access for the neutral (L2) and earth ground terminations. The Flasher alarm includes
an AC solid state timer that switches two 25W halogen lamps at a frequency of about
forty flashes per minute.

2.2 Specifications

Dimensions
(H x W x D )
Local Alarm
pn 1060-024
Remote Alarm
Pn 1060-034
Flasher Alarm
Pn 1060-026
in (cm)
9 x 6.3 x 6.75
(22.9 x 16.0 x 17.1
9 x 6.3 x 6.75
(22.9 x 16.0 x
17.1)
10 x 14x 2
(25.4 x 35.6 x 5.1)
Weight lbs (Kg) 3 lb (1.4) 2.5 lb (1.1) 3 lb (1.1)
Operating
Temperature
F (C)
-32 deg to 122
deg
(0 deg to 50 deg)
-32 deg to 122
deg
(0 deg to 50 deg)
-32 deg to 122
deg
(0 deg to 50 deg)
Relative Humidity
Non condensing
0 to 90%
0 to 90%
0 to 90%