Haier DW-86L386, DW-86L628, DW-86W420, DW-86L388, DW-86L626, DW-86L288 user manual Safety Precautions

Page 5

Optional Accessories

This recorder has been calibrated at the factory. It retains the temperature accuracy and calibration even when there is a power supply interruption. Should there be a need to recalibrate the recorder, please follow the procedures below.

a)Let the unit run continuously until it reaches a stable condition. The recorder has registered temperature for at least two hours.

b)Place a standard thermocouple in a 100 ml container filled with 10% glycerin solution. Place the container and the temperature recorder sensor in the unit. Do not place the recorder sensor in contact with any liquid.

c)After four hours of soak time at the stable freezer temperature, the container temperature should match the freezer temperature. Then compare the thermocouple reading with the temperature recorder reading. If there is a discrepancy, use the button (#1) or the button (#2) on the recorder to move the pen on the chart to reflect the correct temperature.

Note: The ink pen will operate 5 seconds after the button is released.

Caution: The temperature recorder is preset during production, so resetting of the temperature recorder is not recommended unless absolutely necessary.

22

Safety Precautions

This manual and labels attached to the product include important safety precautions required to operate this product. Customers are advised to follow instructions provided to avoid potential damage to product and injury to operators.

Note: Use this product only in the way described in the product literature and this manual. Before using it, verify that this product is suitable for its intended use.

Do not modify system components or use unauthorized parts as this will void the product warranty.

This sign is directly related to operator safety and equipment safety. The contents and instructions that follow the sign must be strictly followed.

This sign refers to prohibited actions. Violation may result in damage to the freezer or compromise the operator's safety.

This label is adhered to a cover that protects a high voltage area to reduce the chance of electric shock. Access to the cover should only be handled by a qualified engineer or technician.

NOTICE!

1.Freezer must be placed at least 200mm away from the wall to allow maximum air circulation.

2.Changing of the Refrigerants should be done according to the Service Manual.

NOTICE!

For the most efficient energy consumption and optimized compressor life,the temperature should be set at the highest temperature suitable for the samples to be stored,normally set at -50 to -80 .

When power to the unit is on, switch on the battery charger. When the unit

switch

battery Charge

 

 

is not in use switch power off and then switch off the battery charger..

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Grounding mark

Wear Cryogloves before opening the door.

3

Image 5
Contents DW-86W420 Ultra Low Temperature Storage Freezer Table of Contents Features of the Haier ULT Freezer Technical ServiceCO2 and LN2 backup cooling system Optional AccessoriesDiagram of parts Description of Key Parts and FunctionsSafety Precautions Proper Usage Replacement of Recorder Battery Temperature recorderPower supply Wiring diagram Refrigeration diagramTechnical data Packing list SpecificationsPre-Installation Information Recycle Rechargeable Battery InstallationLocation for installation EnvironmentInitial Loading Troubleshooting GuideInstallation Operation after a power outage Routine MaintenanceBuzzer alarm resumption after cancellation Display and AlarmsRemote alarm terminals Temperature Adjustment and SettingAlarm Conditions Display and alarm

DW-86L386, DW-86L288, DW-86L626, DW-86L388, DW-86W420 specifications

Haier's line of biomedical freezers, including the DW-86L628, DW-86W420, DW-86L388, DW-86L626, and DW-86L386, showcases the company's commitment to innovation, reliability, and advanced technology in preserving sensitive biological materials. These models are designed to meet the needs of laboratories, hospitals, and research institutions, providing optimal storage solutions for a variety of critical applications.

The DW-86L628 is recognized for its impressive capacity of 628 liters, making it ideal for facilities that require large storage volumes. It features a temperature range of -10°C to -86°C, ensuring that vaccines, biological samples, and pharmaceuticals are stored under precise conditions. With a powerful refrigeration system and an efficient insulation design, this model ensures rapid cooling and maintains consistent temperatures.

The DW-86W420, on the other hand, balances capacity with energy efficiency. This model offers a slightly smaller design, suitable for environments where space is at a premium. It includes a user-friendly digital display that allows for easy monitoring of temperature and system performance, while also providing real-time alerts for any deviations outside of the set parameters. Additionally, it is equipped with a patented eco-friendly refrigerant, reducing the environmental impact while maintaining high performance.

In the DW-86L388, Haier emphasizes robust safety features. This freezer is engineered with a dual-lock system to prevent unauthorized access, alongside an advanced alarm system that notifies users in the event of power failures or temperature fluctuations. This level of security is crucial for laboratories that manage valuable specimens.

The DW-86L626 is tailored for versatility and comes with adjustable shelving and organization options, allowing users to configure the interior to suit their storage needs. Its peak temperature recovery time is remarkably quick, ensuring that samples remain secure even after door openings.

Finally, the DW-86L386 combines compact design with advanced technology. It is aimed at users with limited space, yet it does not compromise on storage efficiency. The unit is equipped with energy-saving modes, reducing operational costs while ensuring optimal performance.

In summary, Haier's DW-86 biomedical freezer models stand out for their high performance, reliability, and thoughtful design tailored for the rigorous demands of biomedical storage. With an emphasis on safety, energy efficiency, and user-friendly features, these freezers are a smart choice for any facility focused on preserving vital biological materials.