Isotemp￿ Laboratory and Chromatography Refrigerators

Page 2 of 12

 

 

Installation

Selecting a Location

Leveling the Unit

Door Handles

Door Removal and Adjustment

Part Number: 104416

Choose a location for the refrigerator that will provide at least three inches of clearance between the cabinet and any adjacent vertical surface at the sides and rear. Appropriate electrical power must be available. Locate the refrigerator within 6 feet of the power outlet so that no extension cord is required.

The refrigerator must be level in order to provide adequate condensation drainage as well as proper door alignment and operation. The refrigerator should be in its final operating location and set so that it is firmly positioned on the floor. There are four leveling screws, one on each corner. . Level the cabinet front to rear and side- to-side using the corner leveling screws. The leveling screws are accessed by removing the base grille, as described below:

1.Remove the lower grille attaching screws.

2.Grasp the grille with both hands.

3.Tilt the lower end of the base grille toward you.

4.Pull grille away from the refrigerator

(Swinging Door Models Only)

Door handles are packed inside each refrigerator. To mount the handle, lift the door gasket behind the two screws on the front of the door. Attach handle with offset away from the cabinet corner and tighten the screws.

(Sliding Door Models Only)

Each door has its own closing spring located at the top of the door track. Each spring is set for proper tension. If adjustment is required:

1.Check for cabinet level (see above).

2.Remove the door(s) by lifting it and sliding about half way open. The roller will fall into a gap in the upper track. While maintaining upward pressure on the door, pull the bottom outward until it clears the bottom track.

3.Adjust the location of the door rollers in the roller brackets (above door) as shown by loosening the lock nut on the back of the roller bracket and moving the roller up or down.

Closing spring

Lock nut on back

Roller

Fig. 1

Image 5
Contents Chromatography Refrigerator Models Service Division Repairs Technical Support Customer Service Table of Contents Visible Loss or Damage Concealed Loss or Damage Packing List Introduction UnpackingInstallation Use ¼ tab length supports on sides and ½’ in back ShelvesPlace shelves at desired height on four shelf supports Electrical ConnectionControl Layout OperationSetting the Temperature Units Select Calibration Offset Temperature ControllerError Codes Manual/Automatic Condenser Fan Fluorescent Lamps Convenience OutletSymptom Trouble Shooting TableCabinet Cleaning Condenser MaintenanceTemperature Range 1 to 12C Convenience Outlet, Chromatography Refrigerators Only Wiring DiagramPart No Rev G